65
OPERATING PROCEDURE: Summary Display
Measurement items to be displayed
3
Press the return
button to return to the measurement screen.
3
Prints the measured results.
When there is no measured data, long press of the print button
(for about 1 second) advances the paper.
4
Right/Left
Switches the eye (with the cursor) between the right and left eyes.
5
Up/
Down
Selects (with the cursor) the upper or lower measurement item.
6
Next page
Displays the next screen page (to show other measured values).
•
The data export and data erasing functions operate when the print
button is pressed as when the
button is pressed on the measurement screen.
•
To erase all the measured data with a single operation, press and hold the erase
button (for
about 1 second) on the measurement screen.
ALL
Selected to erase all the measurement items except for PD.
AR
AR median value, large area measurement value (L.DATA and L.DIFF)
The trial lens data and contact lens conversion value are displayed at the end of
the measured value.
KM
KM median value
PS
PS (pupil size) measured value
CS
CS (corneal size) measured value
PD
Pupillary distance (distance PD, near PD)
UCVA
Uncorrected VA measurement value
SUBJECT
Subjective measurement value (distance corrected VA value, contrast VA value,
glare VA value, subjective refractive error, near addition power, near VA value)
ACC
Accommodation measurement value (AR measurement value MIN/MAX, pupil
diameter MIN/MAX)
RETRO
Retroillumination analysis value (COI.H, COI.A, POI)
•
The measurement items that have not been measured become blank.