81
DEVICE SETTINGS AND MAINTENANCE: Device Parameter Settings
●
PRINT2 (Print setting 2)
●
Sample printout of erroneous data
When the “44. ERROR DATA” parameter is set to “YES”
and the measured results are erroneous, the measured
values are displayed in yellow on the screen.
In addition, the printed measured data is preceded by
“Err”.
40. SUMMARY
(Summary display)
YES / NO
Display of the summary screen
When "YES" is selected, the summary
button to display the summary
screen is displayed instead of the print
button.
Parameter option
Setting contents
41. AR PRINT
(Print format of AR data)
ALL / SHORT
Selects the print format of AR data.
ALL
: All data and median values are printed.
SHORT
: Only median values are printed.
42. KM PRINT
(Print format of KM data)
ALL / ALL(KM) / SHORT
Selects the print format of KM data.
ALL
: All data and median values are printed.
ALL(KM)
: In KM measurement mode, all data and median values are printed. In
other mode, only median values are printed.
SHORT
: Only median values are printed.
43. CONF. INDEX
(Printing of confidence
index)
YES / NO
Sets whether to print the confidence index.
When “NO” is selected, the confidence index is not displayed on the measure-
ment screen, either.
44. ERROR DATA
(Erroneous data)
YES / NO
Sets whether to display and print erroneous data of AR measurement.
Parameter option
Setting contents
• When the “43. CONF. INDEX” parameter is set to “YES”, “E” is printed out as a confidence index.
45. CAT MARK
(Cataract indication)
YES / NO
Sets whether to print “*” representing that measurement has been performed in cat-
aract measurement mode.
46. ERROR PRINT
(Error print)
YES / NO
Selects whether to print failed AR data.
47. L. DATA PRINT
(AR large area mea-
sured data)
DATA / NO / DIFF / DATA&DIFF
Sets the print contents of large area measured values for AR measurement.
DATA
: The large area measured values (L. DATA) are printed.
NO
: Only the central measured values (normal AR values) are printed. (Large
area measured values are not printed.)
DIFF
: The differentiation (L. DIFF) between the central measured values (normal
AR values) and large area measured values is printed.
DATA&DIFF
: Both the contents of "DATA" and "DIFF" are printed.
Содержание ARK-1S
Страница 1: ...AUTO REF KERATOMETER AUTO REF KERATOMETER ARK 1s ARK 1s OPERATOR S MANUAL OPERATOR S MANUAL ...
Страница 6: ...IV ...
Страница 76: ...70 OPERATING PROCEDURE Operation when Peripheral Devices are Connected ...
Страница 112: ...106 SPECIFICATIONS AND TECHNICAL INFORMATION EMC Electromagnetic Compatibility ...