48
OPERATING PROCEDURE: Retroillumination Image Observation
When analysis is complete, the opacity indexes for the center
(COI. H, COI. A) and periphery (POI) and a circle indicating
the 3 mm range in diameter are displayed.
If the pupil of the image to be captured cannot be detected,
“PS+OVER” or “PS-OVER” is displayed. Capture the image again.
When the pupil size is 3 mm or less in diameter, “----” is displayed in the opacity index POI for the
periphery.
Pressing the start button before image capture returns to the Step 1 display.
3
Switch the eye to be measured and perform retroillumination image observation for
the other eye in the same manner.
COI. H
Opacity size within a diameter of 3 mm of
the center (vertical diameter): mm
COI. A
Opacity proportion within a diameter of 3
mm of the center: %
POI
Opacity proportion within the entire
periphery: %
Prints the opacity indexes and retroillumination image only. Prints the opacity
indexes and retroillumination images of the right eye and left eye when both eyes
have been captured.
Return
Returns to the AR (AR/KM) measurement screen.
Scale
Shows or hides the angle scale and
corneal cylinder axis.
Corneal cylinder axis
Red: Steepest meridian
Blue: Flattest meridian
When KM-measured values have
not been obtained or the center of
the retroillumination image cannot
be detected, the scale
icon is
not displayed.
• To print the retroillumination image on the normal measured results, set the “57. RETRO IMAGE
PRINT” parameter.
ᵏ
5(752,//80,1$7,21
&2,+PP
&2,$
32,