4. Measurement
NIC-TD-0000025-00
- 24
–
<
EMP-2+WLE
>
Confirm the additional following items on
”EMP-2”.
(1)
WLE connection
Set in
“EMP-2+WLE” with MENU screen.
(Refer to 5.5 "WLE connection".)
(2)
WLE solution replacement cycle time
Check whether the optimal cycle time of the WLE solution replacement.
(Refer to 5.2.9 "WLE solution replacement cycle".)
(3)
Waiting time during WLE solution replacement
Check the waiting time for the EMP-2 during WLE solution replacement.
(Refer to 5.2.10 "WLE solution replacement time".)
※
If you use the default setting, confirmation is not required.
If you change the setting time, ensure waiting time is always longer than the time
you have set time of WLE.
(4)
WLE solution injection
After power-up, you must inject WLE solution in advance.
When you press the start-stop button for 1 to 3 sec, WLE solution injection starts.
※
If you press the START button without having to change WLE solution, a warning is
displayed and the measurement does not start.
<
WLE solution injection
>
<
Warning
>