5. MENU screen
- 43 -
5.2.11. Measurement cycle time in static mode
Setting the measurement cycle mode in static mode.
It is set to 50 sec before shipping.
Use the "Static Mode" button to change the setting.
5.3.
Confirmation of measured data (Data Check)
This function should be used to confirm the results of measurements made at a site.
Select "MENU" - "Data Check" on the READY screen to change to the Data Check screen.
*
The concentration unit of measured data is always "μg/m
3
".
*
All measurement results are saved on the SD card. Detailed data can therefore be
confirmed using the SD card. (Refer to 6 "Measured data".)
The latest 10 lots of data are displayed on the Data Check screen.
The latest data is displayed from right-to-left order on a graph and the measurement No. are
shown under the graph to identify each piece of data. (Refer to Figure
below.)
The bar corresponding to each piece of data indicates the average value over the
measurement time, the value of which is displayed on top of the bar. (Refer to Figure
below.)
If an alarm-generating level was set, it is used as the full-scale value of the graph, thus
allowing the measurement results to be compared with the alarm-generating level (red line).
If an alarm-generating level was not set the maximum value of the 10 average values is
used as the full-scale value.