5. MENU screen
NIC-TD-0000025-00
- 46
–
No. Name
Function
①
Serial No.
Displays EMP-2 Serial number.
②
Soft Ver.
Displays the software version.
③
Sig [V]
Displays in volts the amount of light passing through the
absorption cell.
Default factory value of 2V@25 °C.
The voltage changes depending on staining in the cell and the
ambient temperature.
④
Ref [V]
Displays the amount of light of the lamp in volts.
Default factory value of 2V@25 °C.
The voltage changes depending on the ambient temperature.
⑤
Battery [V]
Displays the battery voltage.
⑥
F
The numeric value is a sensitivity expressed as a ratio.
This function displays a numeric value after a recalibration
was carried out on the assumption that the default factory
value was 1.
⑦
TEMP
Displays the temperature inside the cell.
⑧
BUZZER
The buzzer function can be verified by turning this button ON
or OFF.
⑨
VALVE
The valve can be opened and closed by turning this button
ON or OFF.
⑩
PUMP
The operation of the pump can be verified by turning this
button ON or OFF.
⑪
Sound Button
Used to specify whether or not to sound a beep sound when a
button on the LCD screen is pressed.
⑫
SD
The status of the SD card can be verified (Inserted / Not
inserted / Locked).
⑬
Total Time
Displays the power ON time of the equipment.
⑭
Lamp
Displays the power ON time of the lamp.
⑮
Pump
Displays the power ON time of the pump.
CAUTION
Use of a badly deteriorated pump can increase the
margin of error of a measurement value.
Immediately contact our service representative once the
pump needs to be replaced.
(Refer to 10 "Contact".)
CAUTION
Use of a badly deteriorated lamp can increase the
margin of error of a measurement value.
Immediately contact one of our service representatives
once the lamp needs to be replaced.
(Refer to 9 "Contact".)