5. MENU screen
NIC-TD-0000025-00
- 36
–
5.2.
Setting the measurement conditions (Setting)
Ensure to set the measurement conditions before making a measurement on the Setting
screen.
To change or reset the measuring conditions select "MENU" - "Setting" on the READY
screen.
There are two Setting screens, which can be switched using the "Next" button.
エラー
!
<
Setting(1/3)
>
<
Setting(2/3)
>
<
Setting(3/3)
>
No. Name
Function
①
Full Scale button
Used to set the full-scale value of the level meter.
②
Warning Level button
Used to set the alarm-generating level of the display
screen.
③
Total Hg button
Used to clear the amount of mercury adhering to the
zero filter.
④
Date & Time button
Used to set the current time and date.
⑤
Unit button
Used to convert the unit of the measured
concentration.
⑥
Measurement Mode button
Used to set the measurement mode.
⑦
Sig/Ref button
Used to regulate the Sig/Ref Voltage.
⑧
Normalization Temp
Used to set the normalization temperature.
⑨
WLE cycle
setting the time interval between WLE solution
replacement.
⑩
Replacing time
setting the waiting time of EMP-2 during WLE solution
replacement.
⑪
Static Mode
setting the measurement cycle time in static mode.
①
③
⑤
⑥
④
②
⑦
⑧
⑨
⑩
⑪