4.1 Operation example in constant voltage mode
HVA4321
4-7
b)
Monitor signal output
Connect a measuring instrument such as oscilloscope or voltmeter to VOLTAGE MONITOR or
CURRENT MONITOR as required.
Figure4-11. Connection of monitor
c)
Connection of load
Be sure to use the accompanying cable for the cable connected to Hi. To minimize effect from stray
capacitance, shorten the cable length as much as possible and keep it away from the floor or other devices.
Connect the output return wire to TOTAL CURRENT terminal in total mode.
Connect it to LOAD CURRENT in load mode.
Figure4-12. Connection of output cable
電圧計
SI GNAL ON/ OFF STATUS
OUTPUT
MONI TOR
VOLTAGE
CURRENT
LOAD
CURRENT
TOTAL
CURRENT
OUTPUT Lo
オシロスコ ープ
Voltmeter
Oscilloscope
LOAD
CURRENT
TOTAL
CURRENT
OUTPUT Lo
±
10kV MAX
±
10mA MAX
OUTPUT Hi
緩まないよう にし っ かり と
締めてく ださ い。
負荷
附属の高圧ケーブル
附属のリ タ ーンケーブル
Tighten it securely
Accompanying high
voltage cable
Accompanying return cable
Load
Содержание HVA4321
Страница 1: ...INSTRUCTION MANUAL NF Corporation HVA4321 10kV AC DC AMPLIFIER HVA4321 10kV AC DC AMPLIFIER...
Страница 2: ......
Страница 4: ......
Страница 8: ...HVA4321 iv...
Страница 16: ...Figures HVA4321 xii...
Страница 18: ...HVA4321 1 2...
Страница 24: ......
Страница 30: ...2 3 Grounding and power supply connection HVA4321 2 6...
Страница 32: ......
Страница 50: ......
Страница 52: ......
Страница 54: ......
Страница 56: ......
Страница 116: ......
Страница 127: ...6 Troubleshooting 6 1 When suspecting a fault 6 1...
Страница 128: ......
Страница 132: ......
Страница 138: ......
Страница 148: ......
Страница 150: ......
Страница 151: ......