3.4 Power supply items
HVA4321
3-17
!
!
3.4 Power supply items
POWER
It is the power switch of the unit.
LINE
It is connected to single phase AC100V.
Be sure to connect it to the outlet with ground with an accompanying 3 pin power cable.
It is terminal for grounding of casing. If grounding is not available with a power cable, ground this
terminal with a wire of 2 mm
2
or more.
Be sure to ground this unit for safety. Otherwise, a serious accident such as electric shock may occur.
FUSE
Use a 10A/250V fuse with time lag.
When a fuse is replaced, be sure to disconnect the power cable from the outlet. Replace it with the same
standard one.
For this product, the time lag type fuse with the rating of 10A/250V and
φ
2 x 20 mm is used.
WARNING
WARNING