4.1 Operation example in constant voltage mode
HVA4321
4-2
Figure4-2. Constant voltage mode (load mode)
lamp
: Output current of high voltage amplifier
lint
: Leak current with internal capacity
lout
: Output current
lstr
: Leak current with wire capacity
lload
: Load current
lac, ldc
: DC and AC of detected current
In load mode, only load current is directly detected. The current monitor can monitor load current only.
In load mode of constant voltage mode:
○
Advantage
・
Only load current can be monitored.
○
Disadvantage
・
One end of load cannot be directly grounded. There is a current detection resistance
(33
Ω
) between load end and ground.
・
The voltage of output current x current detection resistance (33
Ω
) for set output voltage is an
error in load. (When the load current is 10 mA, the error is 330 mV.)
内部 容量
出力端 子
配線の 容量
負 荷抵抗
電 流検 出抵抗
電 流検 出抵抗
(ト ータ ル用)
( ロー ド用)
電 流モニ タ
OUTPUT Hi
OUTPUT Lo
TOTAL
LOAD
I out
Ii nt
Istr
Il oad
I amp
I dc+I ac
電圧 モニタ
Output terminal
Load
resistance
Wire capacity
Internal capacity
Current detection
resistance
(for total)
Current monitor
Voltage monitor
Current detection resistance
(for load)
Содержание HVA4321
Страница 1: ...INSTRUCTION MANUAL NF Corporation HVA4321 10kV AC DC AMPLIFIER HVA4321 10kV AC DC AMPLIFIER...
Страница 2: ......
Страница 4: ......
Страница 8: ...HVA4321 iv...
Страница 16: ...Figures HVA4321 xii...
Страница 18: ...HVA4321 1 2...
Страница 24: ......
Страница 30: ...2 3 Grounding and power supply connection HVA4321 2 6...
Страница 32: ......
Страница 50: ......
Страница 52: ......
Страница 54: ......
Страница 56: ......
Страница 116: ......
Страница 127: ...6 Troubleshooting 6 1 When suspecting a fault 6 1...
Страница 128: ......
Страница 132: ......
Страница 138: ......
Страница 148: ......
Страница 150: ......
Страница 151: ......