3.3 Monitor and control signal system
HVA4321
3-12
3.3.2
Control signal system
ON/OFF CONTROL
It is the remote/local setting switch. In LOCAL, the unit can be turned ON/OFF with the panel switch.
In REMOTE, the unit can be turned ON/OFF with an external signal. In REMOTE, the operation of
HIGH VOLTAGE POWER SUPPLY ON/OFF can be made with the panel.
HIGH VOLTAGE POWER SUPPLY ON/OFF
Internal DC high voltage ON/OFF
SIGNAL ON/OFF
High voltage amplifier input ON/OFF
OSC INPUT ON/OFF
External oscillator signal ON/OFF
INT DC BIAS ON/OFF
Internal DC bias signal ON/OFF
EXT DC BIAS ON/OFF
External DC bias signal ON/OFF
Figure3-17. Remote/local settings
Change REMOTE/LOCAL when the signal is OFF. If REMOTE/LOCAL is changed with the signal
ON, the signal is forcibly turned OFF.
REMOTE CONTROL INPUT
It is the BNC connector to give ON/OFF control signal input when ON/OFF CONTROL is set to
REMOTE.
Figure3-18. Remote control input
I NT DC BI AS ON/ OFF
SI GNAL ON/ OFF
OSCI NPUT ON/ OFF
EXT DC BI ASON/ OFF
HI GH VOLTAGE
POWER SUPPLY ON/ OFF
TTL(Lo: ON)
TTL(Lo: ON)
TTL(Lo: ON)
TTL(Lo: ON)
TTL( Hi: OFF)
REMOTE CONTROL I NPUT
ON/ OFF
CONTROL
REMOTE
LOCAL
Содержание HVA4321
Страница 1: ...INSTRUCTION MANUAL NF Corporation HVA4321 10kV AC DC AMPLIFIER HVA4321 10kV AC DC AMPLIFIER...
Страница 2: ......
Страница 4: ......
Страница 8: ...HVA4321 iv...
Страница 16: ...Figures HVA4321 xii...
Страница 18: ...HVA4321 1 2...
Страница 24: ......
Страница 30: ...2 3 Grounding and power supply connection HVA4321 2 6...
Страница 32: ......
Страница 50: ......
Страница 52: ......
Страница 54: ......
Страница 56: ......
Страница 116: ......
Страница 127: ...6 Troubleshooting 6 1 When suspecting a fault 6 1...
Страница 128: ......
Страница 132: ......
Страница 138: ......
Страница 148: ......
Страница 150: ......
Страница 151: ......