7.4 Performance test
HVA4321
7-4
7.4.2
Test of constant current mode
The gain accuracy and the monitor accuracy when the standard signal input is given in constant current
mode are checked.
Given voltage input from the standard DC power supply to the external DC bias signal input. Make sure
that the output is +10 mA for +10V input.
Make sure that the error between the actual output and the value of the output monitor is 0.5% or less.
Figure7-3. Check method of constant current mode
HVA4321
直流基準電源
電圧計( 出力モニタ )
電圧計( 出力)
電流検出抵抗
DC standard power supply
Current detection resistor
Voltmeter
(output monitor)
Voltmeter
(output)