5.6 Description of Individual Commands
CA5350
5-19
U1
– Queries of Error Status
Description: Output the error status.
Response message:
C A 5 3 5 0
a
b
c
d
e
f
g
h
i
j
k
< t e r m >
It is composed of the product model and 10 characters constituted by ‘
1
’ or “
0
”. The value
of
a
~
j
in the above response message means the state (‘
0
’ or ‘
1
’) of the corresponding
error.
If any of the following errors happened, the corresponding value will be changed to ‘
1
’,
otherwise, ‘
0
’.
a
: Inappropriate command received
b
: Command containing inappropriate parameter received
c
: Not used
d
: Failed in self-check
e
: Current suppression settings are inappropriate (Current value exceeding the
range was set)
f
: Input current exceeds the range of current suppression
g
: Auto current suppression performed when Zero Check was ON
h
: Not used Always ‘
0
’
I
: Error in the setting memory (NVRAM)
J
: Over input was detected
K
: Not used Always ‘
0
’
Remarks:
Meanings of numerical value match with the corresponding remote
control parameters. Refer to the description of each command for more
details.
U2
– Queries of Bias Voltage
Description: Output the bias voltage value.
Response message:
V
+
0 . 0 0 0 0
< t e r m >
Remarks:
Voltage value read is the value set using
V
command or on the panel
irrespective of whether bias is ON/OFF.
Message terminator,
EOI
Message terminator
,
EOI
Header indicating voltage bias '
V
' fixed
Voltage bias value 4 digits after the decimal point
Polarity '
+
' or '
-
'
Model name