5.1 Preparations Before Use
CA5350
5-5
5.1.3.4 Basic Specifications of GPIB
Interface functions
SH1 Source Handshake full functions provided
AH1 Acceptor Handshake full functions provided
T6
Basic Talker, Serial Poll, and talker cancel function by listener-addressed
provided
Talk-Only function not provided
L4
Basic Listener function and listener cancel function by talker-addressed provided
Listen-Only function not provided
SR1 Service Request full functions provided
RL1 Remote Local full functions provided
PP0 Parallel Poll function not provided
DC1 Device Clear full functions provided
DT0 Device Trigger function not provided
C0
Controller function not provided
E1
Open collector drive
TE0 Extended talker function not provided
LE0 Extended listener function not provided
5.1.4 Precautions on Communication
Input buffer
Commands received are first stored in the input buffer, and then interpreted.
Input buffer size is 128 bytes. Ensure that do not send program message exceeding this
size in one time.
When an inappropriate command is received, it will result in error and program
messages will be stopped executing before receiving
X
Command.