5.6 Description of Individual Commands
CA5350
5-16
S
– Current Suppression Range and Value Setting
Description: Settings of range and current value of current suppression
Setting:
Sv,0
Auto range
Sv,1
±8nA range, 1pA resolution
Sv,2
±80nA range, 10pA resolution
Sv,3
±800nA range, 100pA resolution
Sv,4
±8µA range, 1nA resolution
Sv,5
±80µA range, 10nA resolution
Sv,6
±800µA range, 100nA resolution (factory default setting S0,6)
S,10
Abort running auto range
Remarks:
The firs parameter v indicates current value (unit: A), while second
parameter (0 ~ 6) indicates range to be set.
Setting a value that out of the range to v will result in error.
Setting any number except the above to the second parameter will result
in error.
Setting to auto range
(Sv,0)
will ignore the range setting
(Sv,1 ~ Sv,6)
of
current suppression until auto range is released
(S,10)
, and the optimum
range having maximum resolutions will be automatically found and set.
Current suppression range and value will be same as the time when Auto
Current Suppression was aborted.
T
– Filter Rise Time Setting
Description:
Filter rise time setting
Setting:
T.
1µs
T/
3µs
T0
10µs (factory default setting)
T1
30µs
T2
100µs
T3
300µs
T4
1ms
T5
3ms
T6
10ms
T7
30ms
T8
100ms
T9
300ms
Remarks:
Filter is set to ON/OFF with
P
Command.