5.6 Description of Individual Commands
CA5350
5-18
U0
– Queries of CA5350 Setting
Description: Output for the
CA5350
setting.
Response message:
C A 5 3 5 0
A 0
B 0
C 0
H 2 1
I0
J 0
K 0
M 0 0 1
N 2
P 1
R 0 9
S 0
5
T .
W 0
Y 3
Z 0
< t e r m >
(1) Device model
"CA5350"
Fixed
(2) LCD backlight brightness
Refer to
A
command
(3) Bias voltage ON, OFF
Refer to
B
command
(4) Zero Check ON, OFF
Refer to
C
command
(5) Key last pressed
Refer to
H
command
(6) Input connector Front, Rear
Refer to
I
command
(7) Self-check results
0 : Passed the check
1 : ROM error
2 : RAM error
3 : ROM error and RAM error
4 : NVRAM error
(8) EOI and
X
Command Hold OFF Setting Refer to
K
command
(9)
SRQ
mask
Refer to
M
command
(10) Current suppression ON, OFF
Refer to
N
command
(11) Filter ON, OFF
Refer to
P
command
(12) I/V gain
Refer to
R
command
(13) Auto Current Suppression
0 : Auto OFF (manual)
1 : Auto ON
(14) Current Suppression range
Refer to
S
command
When in Auto mode, range that is actually set will be output.
(15) Filter setting
Refer to
T
command
(16) Output amplifier gain setting
Refer to
W
command
(17) Message terminator
Refer to
Y
command
(18) Auto filter ON, OFF
Refer to
Z
command
Remarks:
Meanings of numerical value match with the corresponding remote
control parameters. Refer to the description of each command for more
details.
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
(16)
(17)
(19)
(18)
Содержание CA5350