3.2 Display at Power
“ON” and Initial Setting
CA5350
3-5
3.2.2 Initialization
CA5350
is reset to the factory default settings in the following cases:
At the time of shipment from factory
All settings are initial values.
When setting in memory number 0 is recalled
In setting memory No. 0, factory default settings are stored. However, the contents of
setting memory No.1 ~ No.9 will not changed.
When an abnormal error is found in the backup settings at power-on
All setting memories (memory No. 1 ~ 9) are reset to the factory default state.
Table 3-1 Setting items and initial values list
Setting items
Parameter range
Initial values
<LCD adjustment>
Backlight brightness
0, 1, 2
2
< Input select>
Input select
FRONT, REAR
FRONT
<Gains>
I/V Gain [V/A]
10k, 100k, 1M, 10M, 100M,
1G, 10G
10k
×10 Gain
×1, ×10
×1
<Filter>
Filter
ON
,
OFF
ON
Auto filter
AUTO
,
MANUAL
AUTO
Rise time
1µs, 3µs, 10µs, 30µs,
100µs, 300µs, 1ms, 3ms,
10ms, 30ms, 100ms, 300ms
10µs
< Current suppression>
Current suppression
ON
,
OFF
OFF
Range
±8nA, ±80nA, ±800nA,
±8µA, ±80µA, ±800µA
±800µA
Current
±(0
~
800.0µA)
000.0µA
<Voltage bias>
Voltage bias
ON
,
OFF
OFF
Bias voltage
-8.000
~
+8.000V
0.000V
<Zero check>
Zero check
ON, OFF
ON
<Remote control>
GPIB address
0
~
30
2