32
CO28IR CO
2
Incubators with HEPA M1271-0051
User’s Manual
Symptom
Possible Cause(s) & Solution(s)
CO
2
LEVEL CONTROL (continued)
Overshoots setpoint but stabilizes;
display
and Fyrite match
(continued)
8. Gas tank regulator set wrong: set for 20 psi.
9. Incubator too heavily loaded: remove some
samples.
10. Incubator being operated without shelving.
11. CO
2
attached to sample port instead of “CO
2
IN”
fitting.
12. If display and Fyrite do not match, see Symptom
“Display and Fyrite reading do not match”
below.
Overshoots setpoint and continues to rise;
display and Fyrite match
1. If injection light goes off at setpoint, problem is a
sticking solenoid valve: call Customer service
for directions.
2. If injection light stays on past setpoint, the
control system needs to be fixed.
3. Inlet pressure should not be more than 20 psi.
4. Debris in solenoid causing it to leak
continuously.
5. Solenoid failed while open.
6. Unit miswired.
7. Controller output failed or shorted.
8. CO
2
sensor or interface failed.
9. CO
2
sensor plugged by debris or condensation.
Rises very slowly
1. Restrictor partially plugged.
2. Filter partially plugged.
3. CO
2
tank regulator set too low.
4. Hose kinked.
5. Poor glass door seal.
6. Hose leak.
7. CO
2
tank contains mixed gas, not 100% CO
2
.
8. May be old version of controller, needs update
to “fast rise” version: call Customer service to
confirm.
Never rises
1. CO
2
tank empty.
2. Solenoid failed while closed.
3. CO
2
controller output failed while open.
4. Restrictor plugged.
5. CO
2
hose kinked.
6. Filter plugged.
7. Setpoint at zero or not set high enough.
8. Sensor has come unplugged: check that the
wires between the sensor and the liner are not
too tight, causing disconnection or intermittent
connection.
Display and Fyrite reading do not match
1. Calibration error: clear chamber for 12 hours
and confirm at "0"
2. Turn setpoint up and down to see if solenoid
works by feeling and listening to valve.
...continued...
Содержание CO28IR
Страница 2: ...ii CO28IR CO2 Incubator with HEPA M1271 0051 User s Manual...
Страница 4: ...iv CO28IR CO2 Incubator with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 6: ...vi CO28IR CO2 Incubator with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 8: ...viii CO28IR CO2 Incubator with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 10: ...x CO28IR CO2 Incubator with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 14: ...2 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 16: ...4 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 22: ...10 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 34: ...22 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 36: ...24 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 40: ...28 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 50: ...38 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 52: ...40 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...
Страница 60: ...48 CO28IR CO2 Incubators with HEPA M1271 0051 User s Manual THIS PAGE IS INTENTIONALLY BLANK...