Hearing Screening Systems (Technical Manual)
80
The updating is activated by pressing “Update device firmware”
button. Depending
on the manager settings, the dialog box for the password entering can appear on the
screen.
Only the specialists can update the device firmware as far as the device
operation can be violated.
To update the software, the key file is provided which is supplied with
Neuro-Audio-Screen manager
distributive. It detects the device with serial number
which can be updated using the function of firmware update. After the updating run,
the program checks the availability of this key. In case the key is not detected or does
not correspond to the device being updated, the warning message will appear on the
screen (Fig. 62).
Fig. 62. The warning message concerning the key file mismatch to the device.
If such message appears on the screen, address to your dealer and inform of the
serial number of the device (it is given in the text of the warning message in brackets).
If the key was accepted, the dialog box for the updating file selection will appear on
the screen (Fig. 63).
Fig. 63. The selection of the updating file.