Over current
This occurs when the control
system detects an excessive
amount of current in an
actuator channel. This may
be due to a faulty actuator
motor, cables or connections
Check the movement of the
actuator is not obstructed.
If the trip is still present after
above checks have been
made, the Seating Module
may be defective. See
section 3.14.
Overtemp
This occurs when the
actuator circuitry has
become too hot.
Allow the seating module to
cool. If it is frequently
overheating check the
condition of the actuator
motors and connections.
If the trip persists contact
the manufacturer.
DIME error
This occurs when the control
system detects an
identification conflict
between two modules in the
system
If a new module has been
introduced:
disconnect the new module
and reboot the wheelchair, if
no trip is present connect the
new module to the system
and reboot the wheelchair, if
the trip reappears then the
new module must be the
cause of the problem.
If there have been no
additions:
Disconnect one module at a
time and reboot the
wheelchair. If the trip is still
present after the above
checks have been made,
contact the manufacturer.
Memory error
This is a non-specific
memory error that could be
caused by any of the
modules within the system
Check all cables and
connections. Reboot the
wheelchair.
If the trip is still present after
the above checks have been
made, then the Power
Module may be defective.
See section 3.12.
Содержание EVO3
Страница 1: ...SERVICE MANUAL Rev D 2014 10 11...
Страница 2: ......
Страница 4: ......
Страница 36: ...www neatech it 36...
Страница 69: ...www neatech it 69 5 2 Wiring diagram Figure 82...
Страница 70: ...www neatech it 70 Figure 83...
Страница 79: ...www neatech it 79 Figure 87...
Страница 90: ...www neatech it 90 6 1 Dimensions...
Страница 92: ......
Страница 93: ...SERVICE MANUAL EVO3 www neatech it...