Maintenance, Cleaning, & Disposal
To keep the EEG32U headbox system in good working condition, follow a regular schedule of preventive
maintenance. Regular preventive maintenance does not involve access to the interior of the EEG32U
headbox and components. For service problems that require corrective maintenance and/or internal
component service, call the Natus Technical Support at
1-800-303-0306
or
, or contact
your local
XLTEK
representative.
Periodically check cable connections and electrodes for damage and wear. Inspect cables for bent pins.
Replace frayed or worn cables. Also, regularly inspect and clean all system components, including:
•
Connectors and jack ports
•
Headbox and headbox cable
•
Electrodes and accessories
The EEG32U amplifier and its components should not be immersed in water or any other fluid. To clean,
use a damp cloth to wipe all surfaces.
Taking basic care of the system and avoiding extreme physical abuse helps prolong the lifespan of the
headbox.
Recommendations
Disconnect the EEG32U headbox from the computer before wiping. Disconnect all
cables. Use a lint-free cloth. Do
NOT
use cleaners on any system component.
Be careful not to allow any fluid to seep into the internal electronic components of the
headbox or breakout box.
Do
NOT
leave the headbox attached to the computer when transporting the unit.
Do
NOT
autoclave, pressure sterilize, or gas sterilize this amplifier.
Do
NOT
soak or immerse the amplifier in any liquid.
A cleaning solution of 70% isopropyl alcohol is recommended.
Use cleaning solution sparingly. Excessive solution can flow into the amplifier and
cause damage to internal components.
Do NOT use petroleum-based or acetone solutions, or other harsh solvents, to clean
the amplifier.