User and Service Manual
XLTEK EEG32U Amplifier
29
Save the impedance check as part
of the study
Click
End and Start Recording
Interpreting an Impedance Check
An impedance check displays bar graphs that show the impedance of each electrode connection. A
green
bar indicates the reading is below the threshold. A
red
bar indicates the reading is above the
threshold. To set the impedance threshold, click one of the
Threshold
buttons in the
Threshold Group
box on the right side of the
Check Impedance
screen. To read a numerical value for the impedance in
kOhms, double click on the headbox picture on the screen after closing Impedance Check. This brings up
a table that lists the impedances for all channels that were checked.
NOTE:
If the impedance check is run with a
Protocol
, the threshold is determined in
the
Action
settings in the
Protocol
tab of the
Edit > Settings
box. Otherwise, the
threshold is determined by manually clicking the option buttons in the Threshold
section of Impedance Check dialog box.