background image

User and Service Manual 

XLTEK EEG32U Amplifier 

 

IEC 61000-4-2:2008, ed 2.0 

Electromagnetic Compatibility (EMC) Part 4-2: Testing 
and Measurement Techniques - Electrostatic 
Discharge Immunity Test 

IEC 61000-4-3 ed 3.0 with 
A1:2007+A2:2010 

Electromagnetic Compatibility (EMC) Part 4-3: Testing 
and Measurement Techniques - Radiated, Radio-
frequency, Electromagnetic Field Immunity Test 

IEC 61000-4-4:2012, ed 3.0 

Electromagnetic Compatibility (EMC) Part 4-4: Testing 
and Measurement Techniques - Electrical Fast 
Transient/Burst Immunity Test 

IEC 61000-4-5:2014, ed 3.0 

Electromagnetic Compatibility (EMC) Part 4-5: Testing 
and Measurement Techniques - Surge Immunity Test 

IEC 61000-4-6 ed 2.0 with A1:2004 + 
A2:2006 

Electromagnetic Compatibility (EMC) Part 4-6: Testing 
and Measurement Techniques - Immunity to 
Conducted Disturbances, Induced by Radio-frequency 
Fields 

IEC 61000-4-8:2009, ed 2.0 

Electromagnetic Compatibility (EMC) Part 4-8: Testing 
and Measurement Techniques - Power Frequency 
Magnetic Field Immunity Test 

IEC 61000-4-11:2004, ed 2.0 

Electromagnetic Compatibility (EMC) Part 4-11: 
Testing and Measurement Techniques - Voltage Dips, 
Short Interruptions and Voltage Variations Immunity 
Tests 

IEC 61000-3-2:2014, ed 4.0 

Electromagnetic Compatibility (EMC) Part 3-2: Limits - 
Limits for Harmonic Current Emissions 

IEC 61000-3-3:2013, ed 3.0 

Electromagnetic Compatibility (EMC) Part 3-3: Limits - 
Limitation of Voltage Changes, Voltage Fluctuations 
and Flicker in Public Low-voltage Supply Systems 

CISPR 11 ed 5.0 with A1:2010 

Industrial, Scientific and Medical (ISM) Radio-
Frequency Equipment - Electromagnetic Disturbance 
Characteristics - Limits and Methods of Measurement 

 

Содержание XLTEK EEG32U

Страница 1: ...XLTEK EEG32U User and Service Manual...

Страница 2: ...l rights reserved This manual contains proprietary information which is protected by copyright and may not be copied in whole or in part except with the prior written permission of Natus Medical Incor...

Страница 3: ...CC 14 Warnings and Cautions 15 General Warnings 15 Electrical Warnings and Cautions 16 Patient Environment Warnings and Cautions 17 Transportation Warnings 18 Conducted Immunity Warnings 18 Procedures...

Страница 4: ...Isolation Transformer Details 31 Maintenance Cleaning Disposal 32 Recommendations 32 Disposal 33 Troubleshooting 34 Theory of Operation 35 Introduction 35 System Overview 35 Features of the EEG32U Sy...

Страница 5: ...sed in conjunction with Natus NeuroWorks software to acquire scalp electroencephalographic EEG signals The EEG32U Amplifier is intended to be used by trained medical professionals and is designed for...

Страница 6: ...onvention Description Function This symbol denotes a warning or important information that should not be missed Read all warnings and cautions carefully before starting the system for the first time A...

Страница 7: ...0 1Hz 70 Hz bandwidth EEG Channel Hardware Gain 125 Maximum Differential AC Input Before Clipping Referential 20 mV Maximum Operational DC Input Voltage Electrode Offset 1000 mV Input Bias Current 20...

Страница 8: ...USB 2 0 USB 2 0 Full Speed 12Mbit sec Amplifier Mechanical Interface Cable USB 2 0 cable maximum 3m Approximate Size 160 mm l x 97mm w x 29mm d Environmental Conditions for Use Operating Environmental...

Страница 9: ...2011 ANSI AAMI ES60601 1 2005 R 2012 C1 2009 R 2012 and A2 2010 R 2012 IEC 60601 1 2005 C1 2006 and C2 2007 Third Edition CENELEC EN 60601 1 2006 A1 2013 Medical electrical equipment Part 1 General re...

Страница 10: ...bility EMC Part 4 6 Testing and Measurement Techniques Immunity to Conducted Disturbances Induced by Radio frequency Fields IEC 61000 4 8 2009 ed 2 0 Electromagnetic Compatibility EMC Part 4 8 Testing...

Страница 11: ...t Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 Group 1 The EEG32U uses RF energy only for its internal function Therefore its RF emissions are very low and not likely to cause...

Страница 12: ...ould be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Complies Mains power quality should be that of a typical commercial or hospital...

Страница 13: ...the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site1 should be less than the compliance level in each frequency2 Interfe...

Страница 14: ...800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217 Hz 2 0 3 28 1 845 1 970 2 450 2 400 2 570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 2 0 3 28...

Страница 15: ...ted in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interfere...

Страница 16: ...sions or decreased immunity of the equipment and may cause the system to be non compliant with the requirements of IEC 60601 1 2 2007 The computer and the peripherals used with the EEG32U headbox and...

Страница 17: ...han 30 cm 12 inches to any part of the EEG32U System including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result Electrical Warnings and Caut...

Страница 18: ...ied via MPSO with a separating transformer Do NOT connect electrical equipment which has not been supplied as a part of the system to the MPSO Patient Environment Warnings and Cautions NOTE The patien...

Страница 19: ...transports carts or operating surfaces Natus is not responsible for any injury or damage that may result from improper cable storage during transport TIPPING HAZARD During transport the user should g...

Страница 20: ...atic electricity to damage the product components Avoid carpets in cool dry areas If provided leave the product components in their anti static packaging until ready to be installed Take care when con...

Страница 21: ...e is present on the EEG waveforms try to identify possible culprits by disconnecting nearby equipment from the common power source Lay out the interconnection cables as far as possible from the cables...

Страница 22: ...NTION Consult Accompanying Documents and Consult Accompanying Documents Protective Earth Ground Type BF Equipment Dangerous Voltage Alternating Current Direct Current 5VDC 0 2A Power On Power Off EU o...

Страница 23: ...e or Static Sensitive Manufacturer Information Reference Number This is the part number for the device The ETL mark is a safety symbol that shows the product was independently tested and certified to...

Страница 24: ...uct Images EEG32U Headbox EEG32U Headbox 1 Standard 10 20 Channel Inputs 2 Impedance LEDs 3 Patient Ground 4 Additional Referential Inputs 5 Impedance Threshold LEDs 6 Impedance Button 7 Photic Stim C...

Страница 25: ...EEG32U make sure the following items are included EEG32U headbox USB 2 0 cable Patient event switch Optional Accessories EEG accessory kit NOTE The EEG32U amplifier should be used only with cables tr...

Страница 26: ...efer to the Transport System Specifications and Maintenance section for placement details and cautions for the different cart transportation options Refer to the corresponding Instructions for Use for...

Страница 27: ...Patient Event input on the headbox 4 If a Photic Stimulator is needed insert the Photic Stimulator into the Photic Stimulator connection on the base unit 5 When you are ready to run a study connect th...

Страница 28: ...crovolts peak to peak The accuracy of the gain is 1 NOTE For more information on setting up a montage consult the online Help Channel Test While you are working in the NeuroWorks live recording screen...

Страница 29: ...are connected To Activate an Impedance Check Do one of the following On the keyboard press the minus key on the number pad Choose Controls Impedance Check Press the Impedance button on the headbox hol...

Страница 30: ...ne of the Threshold buttons in the Threshold Group box on the right side of the Check Impedance screen To read a numerical value for the impedance in kOhms double click on the headbox picture on the s...

Страница 31: ...on System setup and installation should be performed by Natus qualified personnel only XLTEK Trolley Specifications Mounting Location Default Equipment Equipment Weight lbs Maximum Load lbs including...

Страница 32: ...Natus is not responsible for any injury or damage that may result from inadequate poorly constructed or unapproved transports carts or operating surfaces Natus is not responsible for any injury or da...

Страница 33: ...U amplifier and its components should not be immersed in water or any other fluid To clean use a damp cloth to wipe all surfaces Taking basic care of the system and avoiding extreme physical abuse hel...

Страница 34: ...life when disposing of the EEG32U amplifier and its components it is recommended that federal state and local laws be followed for proper disposal of printed circuit boards plastics and metal parts F...

Страница 35: ...rly into the headbox not loosely Make sure there are no apparent breaks in the patient electrode cables Are any of the electrodes touching If so they are causing a short circuit and will develop an ar...

Страница 36: ...puter Allows selection of reference signal Provides internally generated test signals sine and square for user assessment of equipment performance Components of the Communications Link Digitized wavef...

Страница 37: ...tion the NeuroWorks computer calculates and transmits the channels gains to each digital board Whenever a study starts the computer uses these stored values to scale the digitized signals channel by c...

Страница 38: ...3 Choose Shut Down from the Start menu 4 Select Restart the computer and click Yes Step 4 Shut Down the Computer Sometimes you need to shutdown the computer completely in order to solve a problem 1 C...

Страница 39: ...Service Manual 10440 Natus Photic Stimulator 026274 4 Port USB hub 026275 12VDC power adaptor 105546 EEG32U TROLLEY MOUNT 008556 EEG32U MOUNT 102181 EEG32U Brain Monitor Breakout Box Roll Stand 10351...

Страница 40: ...ons 6 Environmental Conditions for Use 7 Host Interface 7 Modes of Operation 6 Photic Stim Interface 6 User Interface 6 System Images 23 Testing the Amplifier 26 Theory of Operation MERGEFORMAT 35 EEG...

Страница 41: ...C Standards of Compliance and Normative References 8 Safety Standards of Compliance and Normative References 8 Testing Calibration 27 Channel Test 27 Available Signal Settings 28 Signal Control 27 Imp...

Страница 42: ...taffed with technical and clinical experts 24 7 support remote support via WebEx or VPN the largest clinical and technical field support network in Neuro Sleep and customized service contracts that in...

Отзывы: