User and Service Manual
XLTEK EEG32U Amplifier
17
ELECTRICAL SHOCK HAZARD:
Do
NOT
service the system. Refer servicing to
qualified personnel only. Do
NOT
use repaired components without proper testing.
The computer attached to the EEG32U should use a three-wire power cord with a
hospital grade plug. For grounding reliability, only connect the device to a hospital
grade or hospital-only receptacle. Inspect the power cord often for fraying or other
damage. Do
NOT
operate the system with a damaged power cord or plug.
If there is an isolation transformer with the system, do
NOT
place the isolation
transformer on the floor. Only plug XLTEK recommended components into the
isolation transformer.
Do NOT connect non-medical equipment which has been supplied as part of the system
directly to the wall outlet when the system is supplied, via MPSO, with a separating
transformer.
Do NOT connect electrical equipment which has not been supplied as a part of the system to
the MPSO.
Patient Environment Warnings and Cautions
NOTE:
The patient environment is defined as the area within 1.5 meters of the
patient laterally and within 2.5 meters of the floor in the area occupied by the patient.
Connect all patient electrodes to fully electrically isolated physiological devices only.
Connecting patient electrodes to any other device or external outlet may result in
personal injury.
The patient headbox accepts only touch-proof style electrode inputs. Do
NOT
attempt to use any other style of patient electrode input.
To prevent cabling entanglement/strangulation, ensure no loose cables are
accessible to the patient.
No parts of the ME equipment shall be serviced or maintained while in use with a patient.
Patient connections are
NOT
intended for direct cardiac contact.
Do
NOT
touch any earth-grounded components of the EEG32U amplifier and the
patient simultaneously.