User and Service Manual
XLTEK EEG32U Amplifier
15
Warnings and Cautions
General Warnings
NOTE:
It is recommended that all data be stored using redundant storage
capabilities. This can help to minimize data loss in the event of a failure of the
primary drive.
Proper use of this device depends on careful reading of all instructions and labels
that come with or on the system.
The EEG32U headbox is classified as an IPX0
– ordinary degree of protection
against ingress of water according to IEC 529. The EEG32U is not water proof.
Inaccurate measurements may be caused by incorrect application or use.
Only qualified personnel should operate this equipment.
Only use the
EEG32U
amplifier in conjunction with approved devices and
accessories.
Use of cables other than those specified or sold by the manufacturer on the
equipment, may result in increased emissions or decreased immunity of the
equipment and may cause the system to be non-compliant with the requirements of
IEC 60601-1-2:2007
The computer and the peripherals used with the EEG32U headbox and the
NeuroWorks EEG system must either be approved by
XLTEK
and supplied as part
of an IEC 601 approved system or it must be approved to IEC 950 or similar and
kept outside of the patient environment (that is, the area within 1.5 meters of the
patient laterally and within 2.5 meters of the floor in the area occupied by the
patient).
To ensure the validity of signals, do not operate the device near any sources of
electromagnetic interference.
Turn off the system power before cleaning. Prevent detergent solution or cold
sterilization agents from seeping into the electronics of the system. Be careful
around all connectors and edges. Do not use abrasive agents.
The accessories of this device may include several kinds of disposable, sterile
needle electrodes. These needles may be labeled as
STERILE
. The EEG32U is not
sterile and cannot be made sterile.