XLTEK EEG32U Amplifier
User and Service Manual
36
B. Digital Board
•
Analog to Digital converters digitize the amplified EEG signals. Signals are periodically sampled
and converted to a binary number.
•
The micro controller transmits the digitized signals to the computer over USB 2.0 connection.
•
Contains non-volatile storage for calibration values.
•
During factory calibration, the NeuroWorks computer calculates and transmits the channels gains
to each digital board. Whenever a study starts, the computer uses these stored values to scale
the digitized signals, channel by channel.
•
Equipped with a Patient Event connector that detects an Event Switch closure and transmits it to
the computer.
•
Equipped with a Photic stimulator connector that allows to:
•
Transmit commands from the computer to the photic stimulator over USB 2.0 connection.
•
Transmit feedback (flash strobe marks) from the photic stimulator to the computer over
USB 2.0 connection.