OLYMPIC CFM 6000
Operator’s Manual
29
Tightening the Tilt Adjustment Hinge
If the tilt adjustment of the CFM cart is loose — allowing the CFM 6000 to move backward
and forward too easily — the tilt adjustment hinge requires tightening.
Required items:
Allen wrench, small
To tighten the tilt adjustment hinge:
Using the Allen wrench, tighten the screws on the tilt adjustment hinge (see Figure 11).
Figure 11
Tightening the tilt adjustment hinge
Updating Software
Natus Medical Incorporated periodically updates and releases software for the CFM 6000.
To identify the software version on your CFM 6000:
Touch , and then touch
ABOUT CFM
.
The About CFM message box shows the version and build number for the software
currently installed on your CFM 6000.
To update software when version 2.0 or later is already installed:
1
Touch
TOOLS
, touch
SYSTEM
, and then touch
S/W UPDATES
.
2
Touch
OPEN TRAY
to open the CD tray.
3
Place the software disk in the CD tray, touch
CLOSE TRAY
, and then touch
START UPDATE
. The system automatically restarts.
4
After the CFM 6000 restarts, select from the following choices:
INSTALL
. This option removes all system settings and patient sessions, and
reformats the hard disk. Use this option when you have archived all patient
sessions and want to clean off the hard disk. You need to re-enter all custom
settings under
TOOLS
,
SETTINGS
when you use this option.
UPGRADE
. This option retains system settings and patient sessions. However,
it is recommended that you archive all patient sessions before upgrading.
FIELD SERVICE
. This option is to be used only when qualified personnel are
servicing the device in accordance with the Olympic CFM 6000 Service
Manual.
Use an Allen wrench to
tighten the screws on the
tilt adjustment hinge