28
OLYMPIC CFM 6000
Operator’s Manual
Checking Hard Disk Space
If you do not delete files after you archive patient data or do not permanently remove
files from trash, you might need to periodically check the amount of hard disk space
used so that you know when you need to free up space.
To check the amount of free disk space:
Touch
TOOLS
, touch
SYSTEM
, and then touch
HARD DISK
. You can see how much
disk space is in use, and how much space is used by deleted files in trash.
Cleaning and Maintenance
Cleaning the CFM 6000 and the CFM Cart
Required items:
Soft cloth
Mild cleaning disinfectant (for example, Virex
™
Tb, Virustat
®
, Coverage
®
,
Sporicidin
®
cleaners)
To clean the CFM 6000:
1
Turn off the power and unplug the power cord before cleaning the outer surfaces of
the CFM 6000.
2
Using a soft cloth dampened with mild detergent, wipe the exterior of the
CFM 6000, including the touch screen, amplifier module, and power cable.
To clean the CFM Cart:
Using a soft cloth dampened with mild detergent or disinfectant, wipe the CFM cart,
including the handle, foot pedal, and both interior and exterior regions of the drawer.
WARNING
To avoid electrical shock hazard, always turn off the power and unplug the device prior
to cleaning the device.
To avoid damage to internal electronics and wiring, do not allow liquids to enter the
CFM 6000 enclosure or its vents.
CAUTION
Never autoclave the CFM 6000 or its components as doing so will cause damage and
void the warranty.
Only technically qualified personnel should perform service procedures. On request, Natus
Medical Incorporated will provide circuit diagrams, parts lists, and instructions to assist
qualified technical personnel to repair the CFM 6000 to the printed circuit board (PCB) level.
To obtain service, see
To prevent bio-contamination, clean exterior of CFM per hospital guidelines.
NOTICE!
Use of sodium hypochlorite (bleach) may damage the instrument surfaces.
Use only mild cleaning detergents; other cleaning agents may damage the exterior of the
CFM 6000 and the CFM cart.
Dilute cleaning agents according to their label instructions.