10
OLYMPIC CFM 6000
Operator’s Manual
Starting and Stopping the CFM 6000
To start the CFM 6000:
1
Attach the power cord to the rear panel of the device, and then connect it to a
hospital-grade electrical outlet. If the CFM 6000 is on a cart, lock the cart wheels.
2
Connect the amplifier module cord to the CFM 6000.
3
Press the power switch to the
ON
(
|
) position, and wait for the CFM 6000 main
screen to appear.
To shut down the CFM 6000:
1
Press the CFM 6000 power switch to the
OFF
(O) position, and unplug the power
cord from the electrical outlet.
2
Remove the electrodes from the patient and discard the electrodes.
3
Unplug the amplifier module, and clean the device and module according to the
instructions in
Preparing the Patient — Applying Electrodes
Required items:
Electrodes — hydrogel or low-impedance needle
Skin prep gel, such as NuPrep
™
brand (if using hydrogel electrodes) (part no. 62060)
2x2 or 4x4 gauze pads (if using NuPrep skip prep gel)
NOTE
#
It is recommended that you connect the amplifier module to the CFM 6000 before placing
electrodes on the patient.
NOTE
#
If recording is in process, first touch
9
RECORD
to stop recording.
CAUTION
To avoid electrical shock hazard, ensure that the cover is placed over the USB port connector
when not in use, and never touch the connector and patient simultaneously. For additional USB
cautions, see page 3.
NOTE
#
Read the warnings and cautions on page 2 before operating this device.
To obtain assistance or reorder items, see