20
OLYMPIC CFM 6000
Operator’s Manual
Figure 10
EEG Position Indicator
The EEG position indicator shows the following:
The pink area in the position indicator represents the 15 seconds marked by the cursor
in the CFM tracing.
The dark horizontal bar in the position indicator illustrates where the displayed EEG
tracing occurs within the cursor in the CFM tracing.
The length and width of the dark horizontal bar changes as EEG speed and amplitude
are changed. See
Changing EEG Speed and Amplitude
on page 20 for instructions on
changing the speed and amplitude.
Changing EEG Speed and Amplitude
Modify the speed and amplitude to increase or decrease the level of detail displayed in the
EEG tracing.
To change the EEG speed or amplitude:
1
With the EEG tracing displayed, touch
to enlarge the EEG window and display
the available options.
2
Select the desired speed in mm/sec. Increasing the speed decreases the amount of
EEG tracing displayed in the window; decreasing the speed increases the amount of
tracing displayed in the window.
3
Select the desired amplitude in µVolts. Amplitude indicates the upper and lower
boundaries of the tracing. Decreasing the amplitude setting shows more detail;
increasing the amplitude setting shows more major events.
4
Touch
to reduce the EEG window and view the EEG tracing with the modified
settings.
Entering Patient Information
Patient information identifies a recording session, and includes patient name, patient ID,
date of birth, and time of birth. You can enter patient information before, during, or after