MADSEN Alpha OAE+ Screener User Manual
51
Appendix 3.
Technical Specifications
Appendix 3.1
Accessories
Standard accessories and optional accessories may vary from country to country – consult your local distributor.
Standard accessories
Carrying case
Docking station, including power adapter and USB cable
Probe (Cable Approx. length: 150 cm/59 inches)
User Manual
Battery
Cleaning cloth
Optional accessories
Ear tips
Probe tips
Probe tip cleaning tool
Label printer with printer cable
Probe (Approx. length: 150 cm/59 inches)
External battery charger
Appendix 3.2
Measurement techniques
TEOAE
Evaluation method:
Noise-weighted averaging, counting of significant signal peaks
Stimulus:
Non-linear click sequence
Stimulus level:
75 dB(A) ± 5 dB in 2 cc coupler, self-calibration depending on ear canal volume
Click rate:
65-80 clicks per second (randomized)
Input filter:
1 to 4 kHz
Test display:
Averaged waveform, no. of TEOAE peaks, noise level, artifact level, stability level and test progress
Result display:
Averaged waveform, no. of TEOAE peaks and overall Pass/Clear Response or Refer/No Clear Response. On
Refer/No Clear Response also noise level, artifact level and stability level.
DPOAE
Evaluation method:
Noise-weighted phase statistics
Stimulus:
Primary tone pair, F2/F1 = 1.24
Available test frequencies:
F2 = 1.5, 2, 2.5, 3, 3.5, 4, 5 and 6 kHz
Default test frequencies:
F2 = 2, 3, 4 and 5 kHz (Pass/Clear Response at 3 out of 4)
Default test level:
L1/L2 = 65/55 dB SPL
Test display:
DP-Gram, DPOAE level, noise level
Result display:
Overall Pass/Clear Response or Refer/No Clear Response, DP-Gram with DPOAE and noise level.
SNR:
6 - 8 dB
Minimum Amplitude:
- 5 dB