MADSEN Alpha OAE+ Screener User Manual
33
If there is large ear canal volume (due to PE tubes or perforation of the eardrum) it may not be possible for the L1/L2 to reach the required stimulus
levels. Adequate stimulus levels are crucial to achieve an accurate response. The result is Inconclusive, and a message indicates that there was a large
ear canal volume or a poor probe fit.
If noise exceeds acceptable levels for more than half the test time, the result is Inconclusive and a message indicates that the noise was too high.
Incomplete test
If the test was stopped, an error message and detailed results are shown. The result screen is different if the test failed or was aborted during
calibration.