26
5.3.2
Starting the test
1.
Press
Start (
)
next to the test protocol and ear you want to test. Calibration and speaker tests are performed.
2.
Calibration. If the probe does not fit correctly, adjust the position of the probe.
3.
Speaker Test. A speaker test ensures both speakers function correctly. The test starts automatically when calibration is completed.
Note: If an “Incomplete” message appears. Adjust the position of the probe. Press
Start (
)
to restart the test.
Note: If the calibration fails due to large ear canal volume (for example, due to a PE tube in the patient's ear), and if the Manual start option
is enabled in Test Settings, then a double arrow appears on the result screen. Press double-arrow (
) to start the measurement with
maximum calibrated output levels.