Protocol 5 and 6
6 kHz
5 kHz
4 kHz
3 kHz
2 kHz
1.5 kHz
Minimum DP amplitude equals -5 dB.
Protocol 5: a Pass/Clear Response result requires that 4 of the 6 frequencies receive a Pass/Clear Response result. The test stops
automatically if 4 frequencies reach a Pass/Clear Response result or 3 frequencies receive a Refer/No Clear Response result.
Protocol 6: no overall Pass/Clear Response or Refer/No Clear Response result is generated. All frequencies are tested.
Protocol 7
5 kHz
4 kHz
3 kHz
2 kHz
Minimum DP amplitude equals -10 dB.
Protocol 7: a Pass/Clear Response result requires that 3 of the 4 frequencies receive a Pass/Clear Response result. The test stops
automatically if 3 frequencies reach a Pass/Clear Response result or 2 frequencies receive a Refer/No Clear Response result.