MADSEN Alpha OAE+ Screener User Manual
15
2.5.3
Battery status
•
The battery symbol is shown in the top right corner of the display, enabling you to view the status of the battery at all
times.
•
When less than 25% of the battery capacity remains, you should change the battery and recharge it as soon as possible.
•
When the battery level drops below 10%, test results can be viewed, but testing is no longer possible.
Symbol Remaining battery capacity
100 - 75%
The battery is fully charged.
75- 50%
50 - 25%
25 - 10%
The battery should be charged.
10 - 0%
The battery is very low and testing is not possible.
If the necessary minimum voltage should drop, Alpha OAE+ switches off automatically. Recharge or change the battery.
2.5.4
Powering the docking station
The Alpha OAE+ docking station is powered through an adapter from the mains outlet.
To comply with the safety requirements for a medical device system, connection to the docking station must always be done outside the
patient area (min. 1.5 meters/5 feet from the patient).
You will find an adapter plug kit when you unpack Alpha OAE+. The adapter plug kit contains a range of adapter plugs you can choose from
to fit your mains outlet.
Connecting to the mains supply
Select the adapter plug appropriate to your mains outlet and fit it on the adapter.
Connect the adapter to the docking station and the mains supply. The POWER indicator will light up green.
Warning: Electrical equipment must be positioned so that there is easy access to disconnect the power supply from the mains.
Disconnecting from the mains supply
To disconnect Alpha OAE+ docking station from the mains supply, remove the adapter plug from the mains outlet.