2 Preparing the SMARTS II for use
14
SMARTS II
2.5 Sensor area
Metallic stickers in the sensor area can lead to measurement errors, in
particular to an underestimation of the electromagnetic field strength.
⚠
WARNING
Underestimation of the
electromagnetic field
Risk of personal injury or death to persons at
risk
Do not attach metallic stickers to the sensor area
Only attach metallic sticker to metallic surfaces
(i.e. the user interface at the top side of the
SMARTS II, see 2.8 Instrument overview)
2.6 Power supply
NOTICE
A power supply is not included
For external power supply operation, the power
supply unit must be acquired separately
Damaged power supply
If a power supply is damaged, live parts may become accessible and
result in electric shock injuries.
!
Do not use a damaged power supply.
When a power supply is opened, live parts may become accessible and
death or electric shock injuries may occur.
!
It is not planned to conduct repairs on the power supply. In case of
visible defects or malfunctions, the entire device must be replaced.
!
Never open a power supply.