7 Specifications
SMARTS II
35
7.1 Monitor Specifications
Parameter
Specification
Frequency Range (nom.)
2 MHz to 100 GHz
Field Measured
Electric Field,
Sensor Design
Hemispherical coverage, Diode and Thermocouple array
Frequency Sensitivity
+6.0 / -3.0 dB (100 kHz to 2.3 GHz)
+4.5 / -2.5 dB (2.3 to 30 GHz)
+2.5 / -6.0 dB (30 to 50 GHz)
+2.5 / -6.0 dB (50 to 100 GHz, Typical)
Test Frequencies
100 MHz, 4 GHz
Alarm Threshold
10%, 25%, 35%, or 50% of Standard
1
Alarm Indications
Audio, LED
Maximum CW Power Density
3000% of Standard
Maximum Peak Power Density
32 dB above Standard
STATUS signal
(nom.)
Levels
-4 V standby, +4 V alarm
Load Impedance,
min.
100 k
Status Relay (for
external power
operation only)
Type
SPDT Reed Relay
Power rating
3 W @ 250 mA, 28 V
Recorder
Output Level (nom.)
-1 V (full scale)
(start of alarm output @ -900 mV)
Load Impedance,
min.
100 k
Accessories Supplied
Connector -- D-SUB, Backshell -- D-SUB, Battery, Operating Manual, Test Report
7.2 General Specifications
Parameter
Specification
Recommended Functional Test Interval
2 years
Battery Type
Ultralife U9VL volt lithium
External Power Requirements
2,
3
12 V to 24 V, 50 mA
Temperature
Operating
-10
C to +55
C
Non-operating
-20
C to +55
C
Dimensions (H x W x D)
127 mm x 206 mm x 239 mm (5.0” x 8.1” x
9.4”)
Weight (approximate)
0.74 kg / 1.6 lb
Color
White
1
Determined by DIP switch setting inside battery compartment. There is an additional 1 dB uncertainty in the
alarm threshold when the SMARTS II is set to 10% of Standard.
2
A switch inside the battery compartment determines whether the battery or an external supply powers the
monitor.
3
The external power supply must be floating (not grounded).