M
EASUREMENT
MODES
66
Both the Force Modulation mode and the Spreading Resistance mode are
an extension of the Static Force mode.
The procedure for a first Static Force mode measurement is basically the
procedure described for the Dynamic Force mode in chapter
(p.47). The main differences are:
- Use a suitable cantilever type (usually CONTR), and select this type in
the Operating Mode panel.
- Select the Static Force mode in the Operating Mode panel.
- Set the force Set point in the Z-Controller panel.
For a more detailed description of the parameters that you can set, refer to
section
Dynamic Force
In the Dynamic Force mode, changes in the dynamic behaviour of the can-
tilever are detected by measuring changes in its vibration amplitude when
it is excited with a sinusoidal signal with a frequency close to the cantilever’s
free resonance frequency. When the cantilever tip comes close to the sample
surface, this vibration amplitude will generally go down. Thus the set point
is the percentage of the vibration amplitude when the cantilever was far
away from the surface. A small percentage means a big reduction, a large
percentage means a small reduction. To minimize tip/sample wear, the set
point should be made as large as possible. However, when the Set-point
becomes too large, the Z-Controller may not be able to optimally follow the
sample surface, and artifacts may occur due to instabilities in the cantilever
vibration.
The procedure for a first Dynamic Force mode measurement is described
for the Dynamic Force mode in chapter
more information on how the vibration amplitude and frequency are set,
refer to section
Содержание easyScan 2 AFM
Страница 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Страница 7: ...7...
Страница 86: ...AFM THEORY 86 Scanner coordinate system x y...
Страница 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Страница 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Страница 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Страница 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Страница 164: ......