(p.54). The height of the line is determined at the start of each scan
line: First the Z-Controller is turned on. Once the tip position is stable, the
Z-Controller is turned off and the tip is moved away from the sample by
the distance set by the parameter Rel. Tip-Pos.
The Constant Height Imaging mode is mainly useful for EFM and MFM
measurements. For more information on how to do Magnetic Force
Microscopy, refer to technical note
TN00031 Magnetic Force Microscopy
,
which can be downloaded from the support section of the Nanosurf web
site at
.
Rel. Tip-Pos
The distance by which the Tip is moved towards the sample from the posi-
tion that corresponds to the set point. A negative setting will move the tip
away from the sample.