S
CAN
HEAD
CONFIGURATION
107
The cantilever editor dialog
The following properties of a cantilever type are entered in this dialog:
Name of cantilever
The name of the cantilever type. This name appears in the cantilever
browser and the Cantilever type drop-down in the Operating Mode Panel.
Spring constant
The (nominal) spring constant of this cantilever type. This value is used to
calculate the correct force Set point in the operating modes that use the
static force for Z-Control. (section
Resonance frequency
The (nominal) resonance frequency of the cantilever type. This frequency
is used for the calculation of the automatic resonance frequency search
range (section
Scan head configuration
The scan head configuration is used to store all calibration values specific
to a certain scan head. The configuration of each scan head is stored in a
file with the extension ‘.hed’. This file is copied from the software installa-
tion CD on installation of the instrument. When you change the scan head,
you should also load the correct configuration file.
The scan head selector dialog
The scan head selector dialog is opened via the menu ‘Options>Config
Scan Head...’.
Содержание easyScan 2 AFM
Страница 1: ...Operating Instructions easyScan 2 AFM Version 1 6...
Страница 7: ...7...
Страница 86: ...AFM THEORY 86 Scanner coordinate system x y...
Страница 159: ...THE SCRIPT CONFIGURATION DIALOG 159 This page was intentionally left blank...
Страница 160: ...AUTOMATING MEASUREMENT TASKS 160 This page was intentionally left blank...
Страница 161: ...THE SCRIPT CONFIGURATION DIALOG 161 This page was intentionally left blank...
Страница 163: ...163 Window Operating windows Imaging 120 Positioning 115 Report 153 Spectroscopy 128...
Страница 164: ......