Model
2818
Rotor
Fixed angle (not included)
Automatic detection of rotor
Yes
Temperature
-10 ºC to 40 ºC (
±
2 ºC)
Timer
0-9 h 59 min
Memory
5 programs
Acceleration rate
3 levels
Braking rate
4 levels
Other functions
Quick spin, pre-cooling
Motor
Induction, maintenance-free
Noise
< 68 dB
Power
220-230 V, 50 Hz
Dimensions
74x38x38 cm
Weight
60 Kg
3. TECHNICAL SPECIFICATIONS
Maximum speed according to rotor
Centrifugation is a separation technique based on the movement of the particles that are
displaced to the distal end of the rotation axis depending on their shape and mass.
The Relative Centrifugal Force (R.C.F.) is the required force to separate particles. It is
given as some value times that of gravity (g).
The R.C:F. is calculated with the following formula:
1.118·10
-5
is a constant.
r= radius, the horizontal distance in centimetres
from the rotation axis to the bottom of the sample tube.
n= rotation speed expressed in revolutions per minute (r.p.m.).
F.C.R.= 1.118·10-5·r·n2
Code
92818000
92818001
92818002
92818003
92818004
Capacity
12x1.5 mL
24x1.5 mL
12x5 mL
12x10 mL
6x50 mL
Max. speed
18000 rpm
15000 rpm
15000 rpm
12000 rpm
10000 rpm
Max. RCF
21300 g
20600 g
18600 g
13000 g
10500 g
4. THEORETICAL FOUNDATIONS
Page 18
Instruction manual 52818000
Version 1 Jul.-08
Axis of rotation
Radio of draft
Pre-cooling function
This function allows the centrifuge chamber to reach quickly the pre-selected tempera-
ture. For this purpose:
- Once temperature has been set press the pre-cooling function button
(2.16)
.
- The rotor will spin at 3000 rpm and the timer will start counting up. The tem-
perature inside the centrifuge will go down quickly until reaching the pre-selected tem-
perature; once it has been reached, the rotor will go on spinning to keep it.
Note: To allow the rotor itself reach the pre-selected temperature pre-cooling time must
be longer. The larger the rotor is, the longer the time must be.
Note: if temperature decreases slowly, check that the ventilation grilles of the quipment
are completely free of obstructions
To start with the centrifugation protocol, press stop
(2.19)
and then start
(2.18)
.
Programs
The centrifuge 2818 presents a memory to store up to 5 different centrifugation pro-
grams. To save a program, proceed as follows:
Select the parameters of the protocol to be saved.
Press Prog. button of the program selection buttons
(2.14)
Then, press the number in which the program is going to be saved.
To select a program preciously saved, press the corresponding button number and the
different parameters will be displayed.
Rotor detection
Once it is operating, the centrifuge detects the rotor used and shows its corresponding
number (Rn) on display
(2.13)
. If pre-selected speed is not adequate for the rotor used, the
centrifuge will stop automatically and will display the error.
Rotor chart
Number
Descripción
Capacity
Max. speed
(rpm)Max. RCF (g)
R0
Rotor waiting to be detected
R1
Fixed angle rotor 45º
12x1.5 mL
18000
21300
R2
Fixed angle rotor 45º
24x1.5 mL
15000
20600
R3
Fixed angle rotor 32º
6x50 mL
10000
10500
R4
Fixed angle rotor 45º
12x5 mL
15000
18600
R5
Fixed angle rotor 30º
12x10 mL
12000
13000
Version 1 Jul-08
Instruction manual 52818000
Page 23
ENGLISH