II.

USER MODE CALIBRATION LINK

A  new function has been added to the Ultrameter that makes calibration
of the unit while in "User" mode easier, and more repeatable and accurate
than  other  calibration  methods.  It  is  recommended  that  this  calibration
method  be  used  to  provide  the  highest  degree  of  confidence  when  the
Ultrameter is used in "User" mode. 

A.

Calibration of Ultrameter for use in User Mode

1. 

Press the 

 or 

   key to power up the Ultrameter.

2. 

Calibrate the unit using a Standard Solution. Refer to 
CALIBRATION in Operation Manual.

3. 

Place the Ultrameter in User mode. Refer to SOLUTION 
SELECTION in Operation Manual. 

4. 

Verify/Set the calibration link. (See below – Setting User Mode 
Calibration Link).

B.

Setting User Mode Calibration "Link"

The  link  function  sets  or  "links"  the  calibration  gain  factor  of  a  Standard
Solution  to  the  User  solution  mode.  Once  set,  the  "link"  will  stay  intact
with  future  calibrations  unless  the  link  has  been  canceled.  For  more
information  on  canceling  the  User  Mode  Calibration  Link  refer  to  the
section below "Canceling User Mode Calibration Link". Follow the steps
below  to  set  either  the  KCl,  NaCl  or  442  calibration  factor  to  the  User
solution mode.

1. 

Press measurement key desired to be “Linked”, i.e., 

      ,

               or

.

2. 

Place the Ultrameter in User mode. Refer to SOLUTION 
SELECTION in Operation 
Manual for selecting the User 
Mode.

3. 

Press the 

 arrow key until 

the menu "

Linc

" appears. 

Fig. 6.

Figure 6

MR

COND

T D S

R E S

COND

T D S

3

4. 

Press the  

 key. The  

instrument will display "

S E L

and the “

User

” Icon. Fig. 7.

Any additional display of KCl, NaCl or 442 icons indicates a link between
the User solution and the other solution displayed. 

5. 

Press the 

 keys to 

select a Standard Solution to be 
linked to the User mode
calibration constant.
(see Fig. 8, 

User

 linked to 

KCl

).

 

N o t e

:    If  none  of  the  Solution  Selection  icons  are  displayed  (i.e.  KCl,

NaCl or 442) nothing has been linked to User mode.

6. 

Press the  

  key to accept the setting. Pressing any of the 

measurement keys will exit without changing the setting. User  
mode "link" is now complete. The User mode will now use the
calibration gain constant used for the calibration of the Standard 
Solution as outlined above. 

C.

Canceling User Mode Calibration "Link"

The Ultrameter must be in 

User

 linked mode in order to cancel the "link".

Refer to SOLUTION SELECTION in Operation Manual. 

1. 

Press “Linked” measurement key            ,                  or

.

Two solution icons will be shown the left side of display - “

User

and another, i.e., “

KCl

”. 

2. 

Press the 

  

key until the menu "

Linc

" appears. Fig. 5. 

3. 

Press the  

   key, the unit will display "

S E L

" and the “

User

” 

Icon.   

T D S

COND

R E S

4

MR

CAL

MCLR

CAL

MCLR

M S

MR

CAL

MCLR

Figure 7

User

Figure 8

KCl

User

Содержание Ultrameter 6P

Страница 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Страница 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Страница 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Страница 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Страница 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Страница 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Страница 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Страница 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Страница 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Страница 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Страница 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Страница 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Страница 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Страница 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Страница 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Страница 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Страница 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Страница 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Страница 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Страница 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Страница 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Страница 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Страница 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Страница 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Страница 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Страница 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Отзывы: