fluid washes into the conductivity cell.

D.  pH and ORP Practices to Maintain Calibration.

1.  

Keep the sensor wet with Myron L Storage Solution.

2.  

Rinse away caustic solutions immediately after use.

ORP calibration solutions are not only caustic, but 5% is considered very
accurate.  By  using  the  pH  zero  setting  (0  mV  =  7  pH)  for  ORP  and
precision  electronics  for  detection,  the  Ultrameter  delivers  better
accuracy  without  calibration  than  a  simpler  instrument  could  using
calibration solutions. 

XI.

CARE and MAINTENANCE

Ultrameters should be rinsed with clean water after use. Solvents should
be avoided. Shock damage from a fall may cause instrument failure. 

A.  Temperature Extremes

Solutions  in  excess  of  160°F/71°C  should  not  be  placed  in  the  cell  cup
area;  this  may  cause  damage.  The  pH  sensor  may  fracture  if  the
Ultrameter temperature is allowed to go below  0°C (32°F). Care should be
exercised not to exceed rated operating temperature.
Leaving  the  Ultrameter  in  a  vehicle  or  storage  shed  on  a  hot  day  can
easily subject the instrument to over 150°F. This will void the warranty. 

B.  Battery Replacement

Dry  Instrument  THOROUGHLY

.

 

Remove  the  four  (4)  bottom

screws.  Open  instrument  carefully;  it  may  be  necessary  to  rock  the
bottom  slightly  side  to  side  to  release  it  from  the  RS-232  connector.
Carefully  detach  battery  from  circuit  board.  Replace  with  9  volt  alkaline
battery.  Replace  bottom,  ensuring  the  sealing  gasket  is  installed  in  the
groove  of  the  top  half  of  case.  Re-install  screws,  tighten  evenly  and
securely.

N O T E :    Because  of  nonvolatile  EEPROM  circuitry,  all  data  stored  in
memory and all calibration settings are protected even during power loss
or battery replacement.

C.  pH/ORP Sensor Replacement

Order  model  RPR.  When  ordering,  be  sure  to  include  the  model  and
serial  number  of  your  instrument  to  ensure  receiving  the  proper  type.
Complete  installation  instructions  are  provided  with  each  replacement
sensor.

” 

24

D.  Cleaning Sensors

1.  Conductivity/TDS/Resistivity

The  conductivity  cell  cup  should  be  kept  as  clean  as  possible.  Flushing
with  clean  water  following  use  will  prevent  buildup  on  electrodes.
However, if very dirty samples — particularly scaling types — are allowed
to dry in the cell cup, a film will form. This film reduces accuracy.  When
there  are  visible  films  of  oil,  dirt,  or  scale  in  the  cell  cup  or  on  the
electrodes,  use  a  foaming  non-abrasive  household  cleaner.  Rinse  out
the cleaner and your Ultrameter is ready for accurate measurements.

2.  pH/ORP

The  unique  pH/ORP  sensor  in  your  Ultrameter  is  a  nonrefillable
combination type which features a porous liquid junction. It should not be
allowed  to dry out. If it does, the sensor can sometimes be rejuvenated
by  first  cleaning  the  sensor  well  with  a  liquid  spray  cleaner  such  as
Windex™  or  Fantastic™  and  rinsing  well.  Do  not  scrub  or  wipe  the
pH/ORP sensor.

Then use one of the following methods: 
1.  

Pour a HOT salt solution ~60°C (140°F), preferably potassium 
chloride (KCI) solution — HOT tap water with table salt (NaCl) 
will work fine — in the sensor well and allow to cool.  Retest.

      Or
2.  

Pour DI water in the sensor well and allow to stand for no more 
than 4 hours (longer can deplete the reference solution and 
damage the glass bulb).  Retest. 

If neither method is successful, sensor must be replaced.  

"Drifting" can be caused by a film on the pH sensor bulb. Spray a liquid
cleaner such as Windex™ or Fantastic™ into the sensor well to clean it.

The  sensor  bulb  is  very  thin  and  delicate.  Do  not  scrub  or  wipe  the

pH/ORP sensor.

25

Sensor
Body

pH/ORP Sensor
     Top View

ORP 
Electrode

pH Glass 
Electrode

Reference 
Junction 
under Glass 
pH Bulb

Содержание Ultrameter 6P

Страница 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Страница 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Страница 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Страница 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Страница 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Страница 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Страница 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Страница 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Страница 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Страница 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Страница 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Страница 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Страница 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Страница 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Страница 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Страница 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Страница 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Страница 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Страница 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Страница 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Страница 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Страница 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Страница 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Страница 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Страница 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Страница 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Отзывы: