background image

 C.  Operation of the Keys  (See Instrument Illustration on page 1)

1.  Measurement Keys in General

Any of the 5 measurement keys in the upper part of the keypad turns on
the  instrument  in  the  mode  selected.  The  mode  is  shown  at  the  bottom
of the display, and the measurement units appear at the right. Pressing a
measurement key does this even if you are in a calibration sequence and
also serves to abandon a change (ref. Leaving Calibration, pg. 15).

2.  COND, RES and TDS keys

These 3 keys are used with solution in the Conductivity Cell.  
Precautions:

While filling cell cup ensure no air bubbles cling on the cell wall.

If the proper solution is not selected (KCl, NaCl, 442 or USER), 
refer to Why Solution Selection is Available, pg. 11 and 
Procedure to Select a Solution, pg. 12.

a.  COND Key

Solution to be tested is introduced into the conductivity cell and a press 

of  

  displays conductivity with units on the right. On the left is 

shown  the  solution  type  selected  for  conductivity.  An  overrange
condition will show only [- - - -] (ref. Solution Selection, pg. 11).  

b.  RES Key

A press of 

  displays resistivity with units on the right. On the left 

is shown solution type selected for resistivity (ref. Solution Selection, pg.
11).  The range of display of Resistivity is limited to between 10 kilohms
(KΩ ) and 30 megohms (MΩ ). A solution outside that range will only show
[- - - -] in the display.

c.  TDS key

A press of

  displays Total Dissolved Solids with units on the right. 

This  is  a  display  of  the  concentration  of  material  calculated  from
compensated conductivity using the characteristics of a known material.
On  the  left  is  shown  solution  type  selected  for  TDS  (ref.  Solution
Selection, pg. 11).

3.  pH and ORP keys

Measurements are made on solution held in the pH/ORP sensor well (ref.
pH and ORP Measuring, pg. 34). The protective cap is removed and the
sensor  well  is  filled  and  rinsed  with  sample  enough  times  to  completely
replace the storage solution. 
After use, the pH/ORP sensor well must be refilled with Myron L Storage
Solution, and the protective cap reinstalled securely (ref. Maintenance of
the pH/ORP Sensor, pg. 9 and Cleaning pH/ORP Sensors, pg. 25).
8

COND

R E S

T D S

a.  pH key

A press of

  displays pH readings. No units are displayed. 

b.  ORP key

A press of 

  displays Oxidation-Reduction Potential/REDOX reading 

in millivolts. 

4.  CAL/MCLR key

A press of

   allows you to enter the calibration mode while

measuring  conductivity,  TDS  or  pH.  Once  in  CAL  mode,  a  press  of  this
key  accepts  the  new  value.  If  no  more  calibration  options  follow,  the
instrument returns to measuring (ref. Leaving Calibration, pg. 15).

If  

 is held down for about 3 seconds, CAL mode is not entered, but

S E L

”  appears  to  allow  Solution  Selection  (ref.  pg.  11)  with  the  Up  or

Down keys. As in calibration, the CAL key is now an “accept” key.
While  reviewing  stored  records,  the  MCLR  side  of  the  key  is  active  to
allow clearing records (ref. Clearing a Record/Memory Clear, pg. 20).

5.  UP or DOWN keys

While measuring in any parameter, the            or                keys activate the 

Memory Store and Memory Recall functions. 
While  in  CAL  mode,  the  keys  step  or  scroll  the  displayed  value  up  or
down. A  single press steps the display and holding either key scrolls the
value rapidly.
While in Memory Recall, the keys move the display up and down the stack
of records (ref. Memory Recall, pg. 20).

III.  

AFTER USING the ULTRAMETER

A.  Maintenance of the Conductivity Cell

Rinse  out  the  cell  cup  with  clean  water.  Do  not  scrub  the  cell.  For  oily
films, squirt in a foaming non-abrasive cleaner and rinse. Even if a very
active  chemical  discolors  the  electrodes,  this  does  not  affect  the
accuracy; leave it alone (ref. Cleaning Sensors, pg. 25).

B.  Maintenance of the pH/ORP Sensor

The  sensor  well  must  be  kept  wet  with  a  solution.  Before  replacing  the
rubber  cap,  rinse  and  fill  the  sensor  well  with  (in  order  of  preference):
Myron L  Storage Solution, an almost saturated KCl solution, pH 4 buffer

9

O R P

pH

CAL

MCLR

M S

MR

CAL

MCLR

Содержание Ultrameter 6P

Страница 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Страница 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Страница 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Страница 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Страница 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Страница 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Страница 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Страница 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Страница 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Страница 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Страница 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Страница 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Страница 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Страница 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Страница 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Страница 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Страница 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Страница 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Страница 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Страница 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Страница 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Страница 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Страница 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Страница 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Страница 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Страница 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Отзывы: