background image

Another solution would have a different tempco because of its ionization
activity.  And,  that  tempco  may  be  a  little  different  at  a  different
concentration  or  temperature.  This  is  why  the  Ultrameter  uses
mathematically generated models for known salt characteristics that vary
with concentration and temperature. 

B.  Finding the Tempco of an Unknown Solution

One  may  need  to  measure  compensated  conductivity  of  some  solution
unlike any of the 3 standard salts. In order to enter a custom fixed tempco
for  a  limited  measurement  range,  enter  a  specific  value  through  the
“USER”  function.  The  tempco  can  be  determined  by  2  different
methods:

1.

Heat or cool a sample of the solution to 25°C, and measure its 
conductivity. Heat or cool the solution to a typical temperature 
where it is normally measured. After selecting USER function, 
set the tempco to 0 %/°C as in Disabling Temperature 
Compensation, pg. 13 (No compensation). Measure the new 
conductivity and the new temperature. Divide the % decrease 
or increase by the 25°C value. Divide that difference by the 
temperature difference.

2.

Heat or cool a sample of the solution to 25°C, and measure its 
conductivity. Change the temperature to a typical measuring 
temperature. Set the tempco to an expected value as in User 
Programmable Tempco, pg. 12. See if the compensated value 
is the same as the 25°C value. If not, raise or lower the tempco 
and measure again until the 25°C value is read.  

C.  Finding the TDS Ratio of an Unknown Solution

Once  the  effect  of  temperature  is  removed,  the  compensated
conductivity  is  a  function  of  the  concentration  (TDS).  There  is  a  ratio  of
TDS  to  compensated  conductivity  for  any  solution,  which  varies  some
with  concentration.  The  ratio  is  set  during  calibration  in  USER  as  in
section  User  Programmable  Conductivity  to  TDS  Ratio,  pg.  13.  A  truly
unknown  solution  has  to  have  its  TDS  determined  by  evaporation  and
weighing. Then the solution whose TDS is now known can be measured
for conductivity and the ratio calculated. Next time the same solution is to
be measured, the ratio is known.

XVII.   pH and ORP MEASURING

 

A. pH
  1. pH as an Indicator

pH is the measurement of Acidity or Alkalinity of an aqueous solution. It is 
also stated as the Hydrogen Ion activity of a solution. pH measures the 
34

effective, not the total, acidity of a solution. 

A 4% solution of acetic acid (pH 4, vinegar) can be quite palatable, but a
4%  solution  of  sulfuric  acid  (pH  0)  is  a  violent  poison.  pH  provides  the
needed  quantitative  information  by  expressing  the  degree  of  activity  of
an acid or base.

In  a  solution  of  one  known  component,  pH  will  indicate  concentration
indirectly.  However,  very dilute solutions may be very slow reading, just
because the very few ions take time to accumulate.

  2. pH Units

The  acidity  or  alkalinity  of  a  solution  is  a  measurement  of  the  relative
availabilities of hydrogen (H  ) and hydroxide (OH   ) ions. An increase in 
(H  ) ions will increase acidity, while an increase in (OH  ) ions will increase
alkalinity.  The  total  concentration  of  ions  is  fixed  as  a  characteristic  of
water, and balance would be 10     mol/liter (H   ) and (OH  ) ions in a neutral
solution (where pH sensors give 0 voltage).

pH  is  defined  as  the  negative  logarithm  of  hydrogen  ion  concentration.
Where (H  ) concentration falls below 10   , solutions are less acidic than
neutral, and therefore are alkaline. A concentration of 10     mol/liter of (H  )
would have 100 times less (H  ) ions than (OH  ) ions and be called an
alkaline solution of pH 9.  

  3. The pH Sensor 

The active part of the pH sensor is a thin glass surface which is selectively
receptive  to  hydrogen  ions.  Available  hydrogen  ions  in  a  solution  will
accumulate  on  this  surface  and  a  charge  will  build  up  across  the  glass
interface.  The  voltage  can  be  measured  with  a  very  high  impedance
voltmeter circuit; the trick is to connect the voltmeter to solution on each
side.

The  glass  surface  encloses  a  captured  solution  of  potassium  chloride
holding an electrode of silver coated with silver chloride. This is as inert a
connection  as  can  be  made  from  metal  to  an  electrolyte.  It  still  can
produce an offset voltage, but using the same materials to connect to the
solution on the other side of the membrane allows the 2 equal offsets to
cancel.

The  problem  is...the  other  side  of  the  membrane  is  some  test  solution,
not potassium chloride. The outside electrode, also called the Reference
Junction, is  of  the  same  construction  with  a  porous  plug  in  place  of a
glass barrier to allow the junction fluid to contact the test solution without 

35

-

-

+

-

7

+

+

-

-

7

-

9

+

+

+

-

Содержание Ultrameter 6P

Страница 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Страница 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Страница 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Страница 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Страница 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Страница 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Страница 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Страница 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Страница 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Страница 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Страница 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Страница 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Страница 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Страница 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Страница 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Страница 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Страница 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Страница 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Страница 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Страница 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Страница 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Страница 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Страница 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Страница 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Страница 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Страница 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Отзывы: