background image

Cell Check

  For Instruments with Software Version 

2.10 & Later.

In  these  versions,  a 

Cell  Check

  feature  has  been  added  to  further

increase the performance of your instrument. This is especially important
when  in 

R E S

  mode  reading  High  Resistivity  or  Ultrapure  waters.  This

feature,  utilizing  technological  improvements,  knows  when  the
Conductivity Cell cup is dirty and calls it to your attention. You  may then
choose  to  clean  the  Conductivity  Cell  cup  or  ignore  it  by  pressing  the
CAL key. Follow the steps below to perform a 

Cell Check

1. 

Press

          key to power up the unit.

2. 

Verify that the cell cup is empty 
of any solution and "

- - - -

" is 

displayed by the Ultrameter.  If 
a reading other than "

- - - -

" is 

displayed, clean the cell cup 
and repeat steps 1 & 2.  See 
"Cleaning Sensors".

3. 

Press the 

 key until 

"

CELL ch

" appears. Fig. 2. 

4. 

Press the 

 key.  If the cell

is clean, “

Good

” will 

momentarily be displayed. Fig. 3.

5. 

If the 

Cell Check

 has failed 

the display will show 

C E L L

 

cLn

”, Fig. 4, alternating 

with a number such as “

5 3

”, 

Fig. 5, indicating a relative 
amount of contamination or dirt 
in the Conductivity Cell. To 
insure the highest quality 
readings, the Conductivity Cell 
cup should be cleaned before 
measuring High Resistivity 
solutions. Rinse the cup with 
DI water several times to clean,

 

and Shake instrument well to 
remove excess water. 
Repeat steps 1-4.   

2

Figure 2

Figure 4

Figure 5

Figure 3

CAL

MCLR

MR

R E S

I.

ENHANCED HIGH RESISTIVITY MEASUREMENTS

The  resistivity  calculations  in  the  Ultrameter  have  been  improved  for
measuring  waters  greater  than  10  Megohms.  When  the  Ultrameter  is  in
one  of  the  solution  modes  (i.e.  KCl,  NaCl  or  442)  and  the  resistivity
reading is greater than 10 Megohms, the Ultrameter performs automatic
temperature compensation for high purity water.  As such, the maximum
possible  value  that  should  be  displayed  for  water  is  18.2.  It  may  be
possible  to  display  readings  higher  than  18.2  if  the  instrument  is  not
calibrated  or  if  solutions 

other  than  water

  are  being  measured.  To

insure  proper  use  of  the  instrument  in  this  mode,  readings  greater  than
20 Megohms will display "

-  -  -  -

"  indicating  an  over-range  condition.  To

obtain  resistivity  readings  for  solutions 

other  than  water

,  the  User

mode  should  be  selected.  In  User  mode  the  Ultrameter  will  display
resistivity measurements up to 30 Megohms.  
An 

Offset  Zero  Calibration

  feature  was  added  to  software  version

2.03, and must be performed by the user, see below. On all later versions
this function is performed at time of manufacture. 

Cell Check

 feature was added to these later versions. See page 2. 

Offset Zero Calibration 

For Instruments with Software Version 

2 . 0 3

When  performing  measurements  of  waters above  10  Megohms,  the

accuracy of the Ultrameter may be improved by performing an offset zero
calibration. Follow the steps below to perform an offset zero calibration.

1. 

Press

         key to power up the unit.

2. 

Verify that the cell cup is empty of any solution and "

- - - -

" is 

displayed by the Ultrameter.  If a reading other than "

- - - -

" is 

displayed, clean the cell cup and repeat steps 1 & 2.  See 
"Cleaning Sensors".

3. 

Press the 

 key until 

"

C A L 0

" appears. Fig. 1. 

4. 

Press the 

 key.   The 

instrument should momentarily 
display a number of counts, and return to Resistivity mode.

5. 

If the calibration has failed the display will show "

Err

".   If an error 

occurs during this step, the cell cup is probably contaminated. 
Rinse the cup with DI water several times to clean and Repeat 
steps 1-4.  

R E S

CAL

MCLR

MR

Figure 1

1

Содержание Ultrameter 6P

Страница 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Страница 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Страница 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Страница 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Страница 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Страница 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Страница 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Страница 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Страница 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Страница 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Страница 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Страница 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Страница 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Страница 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Страница 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Страница 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Страница 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Страница 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Страница 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Страница 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Страница 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Страница 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Страница 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Страница 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Страница 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Страница 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Отзывы: