background image

significant migration of liquids 
through the plug material. Figure 14 
shows a typical 2 component pair. 
Migration does occur, and this limits 
the lifetime of a pH junction, from 
depletion of solution inside the 
reference junction or from 
contamination. The junction is 
damaged by drying out because
insoluble crystals may form in a 
layer, obstructing contact with test 
solutions. See pH/ORP, pg. 25. 

  4. The Myron L Integral pH Sensor

The sensor in the Ultrameter (figure 15) 
is a single construction in an easily 
replaceable package. The sensor body 
holds an oversize solution supply for long 
life. The reference junction “wick” is 
porous to provide a very stable, low 
permeability interface. It is located 
under the glass pH sensing electrode. 
The construction combines all the best 
features of any pH sensor known.

  5. Sources of Error

The basics are presented in pH/ORP, pg. 25. 

 

a.  Reference Junction

The most common sensor problem will be a clogged junction because a
cell was allowed to dry out. The symptom is a drift in the “zero” setting at 7
pH.  This  is  why  the  Ultrameter  does  not  allow  more  than  1  pH  unit  of
offset during calibration. At that point the junction is unreliable.

    

b. Sensitivity Problems

Sensitivity  is  the  receptiveness  of  the  glass  surface,  which  can  be
diminished  by  a  film  on  the  surface,  or  a  crack  in  the  glass.  These
problems also cause long response time.

   

c. Temperature Compensation

pH  sensor  glass  changes  its  sensitivity  slightly  with  temperature,  so  the
further from pH 7 one is, the more effect will be seen. A pH of 11 at 40°C 
would  be  off  by  0.2  units.  The  Ultrameter  senses  the  cell  temperature
and compensates the reading.  

36

Figure 14

KCl solution

Electrode wire

Glass surface

Junction plug

Electrode wire

H+ ions

Figure 15

Junction plug

Platinum button

H+ ions

Electrode wires

KCl solution

Glass
Sleeve

Glass

 Surface

B. ORP/Oxidation-Reduction Potential/REDOX
  1. ORP as an Indicator

ORP  is  the  measurement  of  the  ratio  of  oxidizing  activity  to  reducing
activity in a solution. It is the potential of a solution to give up electrons
(oxidize other things) or gain electrons (reduce).  

Like  acidity  and  alkalinity,  the  increase  of  one  is  at  the  expense  of  the
other,  so  a  single  voltage  is  called  the  Oxidation-Reduction  Potential,
with  a  positive  voltage  showing,  a  solution  wants  to  steal  electrons
(oxidizing  agent).  Chlorinated  water  will  show  a  positive  ORP  value,  for
instance.

  2. ORP Units

ORP is measured in millivolts, with no correction for solution temperature.
Like pH, it is not a measurement of concentration directly,  but of activity
level.  In  a  solution  of  only  one  active  component,  ORP  does  indicate
concentration.  Also,  as  with  pH,  a  very  dilute  solution  will  take  time  to
accumulate a readable charge. 

  3. The ORP Sensor

An  ORP  sensor  uses  a  small  platinum  surface  to  accumulate  charge
without  reacting  chemically.  That  charge  is  measured  relative  to  the
solution,  so  the  solution  “ground”  voltage  comes  from  a  reference
junction - same as the pH sensor uses.

  4.  The Myron L ORP Sensor 

Figure 15 pg. 36 shows the platinum button in a glass sleeve. The same
reference  is  used  for  both  the  pH  and  the  ORP  sensors.  Both  pH  and
ORP will indicate 0 for a neutral solution. Calibration at zero compensates
for error in the reference junction.  

A zero calibration solution for ORP is not practical, so the Ultrameter uses
the offset value determined during calibration to 7 in pH calibration (pH 7
=  0  mV).  Sensitivity  of  the  ORP  surface  is  fixed,  so  there  is  no  gain
adjustment either.

  5. Sources of Error

The  basics  are  presented  in  pH/ORP,  pg.  25,  because  sources  of  error
are much the same as for pH. The junction side is the same, and though
the platinum surface will not break like the glass pH surface, its protective
glass sleeve can be broken. A surface film will slow the response time and

diminish sensitivity.  It can be cleaned off  with detergent or acid, as with

the pH glass. 

37

Содержание Ultrameter 6P

Страница 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Страница 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Страница 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Страница 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Страница 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Страница 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Страница 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Страница 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Страница 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Страница 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Страница 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Страница 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Страница 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Страница 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Страница 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Страница 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Страница 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Страница 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Страница 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Страница 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Страница 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Страница 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Страница 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Страница 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Страница 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Страница 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Отзывы: