background image

will be the uncalibrated sensor. 

N O T E S :  If  a  wrong  buffer  is  added  (outside  of  6-8  pH),“7 ”  and
BUFFER ” will flash, and the Ultrameter will not adjust.
The  uncalibrated  pH  value  displayed  in  step  4  will  assist  in  determining
the  accuracy  of  the  pH  sensor.  If  the  pH  reading  is  above  8  with  pH  7
buffer solution, the sensor well needs additional rinsing or the pH sensor
is defective and needs to be replaced.

5.

Press              or               until the display reads 7.0.

N O T E :  Attempted calibration of >1 pH point from factory calibration will
cause “FAC ” to appear.  This indicates the need for sensor replacement
(ref. Troubleshooting  pg. 27) or fresh buffer solution. The “FAC”  internal
electronic  calibration  is  not  intended  to  replace  calibration  with  pH
buffers.  It  assumes  an  ideal  pH  sensor.  Each  “FAC”  indicates  a  factory
setting for that calibration step (i.e., 7, acid, base). 

You can press              to accept the preset factory value, or you can 

reduce your variation from factory setting by pressing  

       or                 .

6.

Press            to accept the new value. The pH Zero Calibration is 

now complete. You may continue with pH Gain Calibration or exit 
by pressing any measurement key.

b.  pH Gain Calibration

Important

: Always calibrate or verify your Ultrameter with a pH 7 buffer

solution before adjusting the gain with acid or base buffers, i.e., 4 and/or
10, etc. Either acid or base solution can be used for the 2nd point “Gain”
calibration and then the opposite for the 3rd point. The display will verify
that a buffer is in the sensor well by displaying either “Acd” or “bAS”.

1.

The pH calibration mode is initiated by either completion of the 

pH Zero Calibration, or verifying 7 buffer and pressing the 
twice while in pH measurement mode.

2.

At this point the “

CAL

”, “

BUFFER

” and “

Acd

” or “

b A S

” 

18

CAL

MCLR

M S

MR

M S

MR

CAL

MCLR

CAL

MCLR

annunciators will be lit (see Figures 7 and 8).

N O T E :  If  the  “Acd ”  and  “b A S ”  indicators  are  blinking,  the  unit  is
indicating  an  error  and  needs  either  an  acid  or  base  solution  present  in
the sensor well.

3.

Rinse sensor well 3 times with acid or base buffer solution.

4.

Refill sensor well again with same buffer solution.

5.

Press              or               until display agrees with buffer value.

6.

Press             to accept 2nd point of calibration. Now the display 

shows the next type of buffer to be used. 

Single point Gain Calibration is complete. You may continue for 
the 3rd point of Calibration (2nd Gain) or exit by pressing any 
measurement key. Exiting causes the value accepted for the 
buffer to be used for both acid and base measurements. 
To continue with 3rd point calibration, use basic buffer if acidic 
buffer was used in the 2nd point, or vice-versa. Again, match the 
display to the known buffer value as in step 2 and continue with 
the following steps. 

7.

Repeat steps 3 through 6 using opposite buffer solution.

8.

Press             to accept 3rd point of calibration which ends 

Calibration procedure. Fill sensor well with Myron L Storage 
Solution and replace protective cap.

6.  ORP Calibration

ORP electrodes  rarely  give  false  readings  without  problems  in  the
reference electrode. For this reason, and because calibration solutions 

19

M S

CAL

MCLR

MR

CAL

MCLR

Figure 8

Figure 7

pH

BUFFER

CAL

pH

BUFFER

CAL

Содержание Ultrameter 6P

Страница 1: ...Ultrameter Operation Manual Model 6P 9 00 MYRON L COMPANY...

Страница 2: ...ion factor and conductivity TDS ratio Displayed here Temperature readout USER temperature compensation or conductivity TDS ratio Memory Storage Recall pH Calibration ORP Electrode pH Glass Electrode p...

Страница 3: ...A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Ultrameter only The Myron L Company...

Страница 4: ...ing Temperature Compensation 13 3 User Programmable Conductivity to TDS Ratio 13 4 VI CALIBRATION 14 A Calibration Intervals 14 B Rules for Calibration in the Ultrameter 14 1 Calibration Steps 15 2 Ca...

Страница 5: ...On the left of the Ultrameter s liquid crystal display is shown an indicator of the salt solution characteristic used to model temperature compensation of conductivity and its TDS conversion The indic...

Страница 6: ...sample enough times to completely replace the storage solution After use the pH ORP sensor well must be refilled with Myron L Storage Solution and the protective cap reinstalled securely ref Maintena...

Страница 7: ...pulling up 10 COND RES TDS 2 Rinse sensor well 3 times with sample to be measured Shake out each sample to remove any residual liquid 3 Refill sensor well with sample 4 Press 5 Take reading 6 IMPORTAN...

Страница 8: ...or User 4 Press to accept new solution type E Application of USER Solution Type 1 User Programmable Tempco This feature allows you to change your Ultrameter s temperature compensating factor to anoth...

Страница 9: ...e calibration procedures The becomes an ACCEPT key At each point pressing accepts the new calibration value and steps you to the next adjustment or out of CAL mode if there are no more steps To bypass...

Страница 10: ...442 Figure 5 4 Reloading Factory Calibration Cond or TDS If calibration is suspect or known to be wrong and no standard solution is available the calibration value can be replaced with the original f...

Страница 11: ...sensor well by displaying either Acd or bAS 1 The pH calibration mode is initiated by either completion of the pH Zero Calibration or verifying 7 buffer and pressing the twice while in pH measurement...

Страница 12: ...ve memory recall or allow to automatically turn off C Clearing a Record Memory Clear After recalling a certain record location press to clear that memory This space will be the place for the next memo...

Страница 13: ...checked every 2 weeks to ensure accuracy Measuring some solutions will require more frequent intervals B Calibration Tracking Records To minimize your calibration effort keep records If adjustments yo...

Страница 14: ...cement Order model RPR When ordering be sure to include the model and serial number of your instrument to ensure receiving the proper type Complete installation instructions are provided with each rep...

Страница 15: ...extend the sensor s useful life Samples containing chlorine sulfur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the sensor well with...

Страница 16: ...ef pH ORP Sensor Replacement pg 29 pH readings drift or respond 1 Temporary condition due to memory Clean and rejuvenate sensor ref Cleaning slowly to changes in of solution in pH sensor well for long...

Страница 17: ...sensor It is available in quarts and gallons Order SSQ or SSG D Soft Protective Case Padded Cordura Nylon carrying case features a belt clip for hands free mobility Model UCC Registered trade mark of...

Страница 18: ...the graph below Users wanting to measure natural water based solutions to 1 would have to alter the internal compensation to the more suitable preloaded 442 values or stay close to 25 C Some who have...

Страница 19: ...commonly referenced compounds In the LCD display the solution type being used is shown on the left Besides KCl NaCl and 442 there is the USER choice The benefit of USER is that one may enter the tempe...

Страница 20: ...ty of an aqueous solution It is also stated as the Hydrogen Ion activity of a solution pH measures the 34 effective not the total acidity of a solution A 4 solution of acetic acid pH 4 vinegar can be...

Страница 21: ...P Oxidation Reduction Potential REDOX 1 ORP as an Indicator ORP is the measurement of the ratio of oxidizing activity to reducing activity in a solution It is the potential of a solution to give up el...

Страница 22: ...tential REDOX pg 37 Reduce As in ORP TDS Total Dissolved Solids or the Total Conductive Ions in a solution See Conductivity Conversion to TDS pg 32 Tempco Temperature Compensation See Temperature Comp...

Страница 23: ...Ultrameter Operation Manual Addendum Models 4P 6P Software Versions 2 03 2 10 2 51 Later See page 5 to determine the version of software of your Ultrameter UMMA10 01 WEB...

Страница 24: ...e resistivity reading is greater than 10 Megohms the Ultrameter performs automatic temperature compensation for high purity water As such the maximum possible value that should be displayed for water...

Страница 25: ...e Linked i e or 2 Place the Ultrameter in User mode Refer to SOLUTION SELECTION in Operation Manual for selecting the User Mode 3 Press the arrow key until the menu Linc appears Fig 6 Figure 6 MR COND...

Страница 26: ...ink should be canceled See above Canceling User Mode calibration Link 3 Once a link has been established for User mode the link will apply to all measurement modes using User solution selection i e TD...

Отзывы: