INE-
DZB-712
Portable
Multi-parameter
Analyzer
24
During measurement, meter allows you to setup temperature value by manual, mV and pH value
resolution, adjust zero point potential, view pH measurement parameter
(please refer to chapter
‘View pH Measurement Data’ for details),
re-calibrate pH electrode
(please refer to chapter
‘pH Electrode Calibration’ for details),
store or print measured data, etc…
If user selects
Continuous Measurement Mode (please refer to chapter ‘Setup
Measurement Mode’ for details),
when measurement result stable, user could store current
measurement result by pressing ‘STORE’(F1). If USB is connected with computer, meter will
print current measured data.
If user selects
Timed Measurement Mode (please refer to chapter ‘Setup Measurement
Mode’ for details),
meter will automatically measure, calculate and display measurement
result. When set time interval reaches, meter will store measurement result automatically; if
USB is connected with computer, meter will print current measured data and automatically
start next measurement.
If user selects
Balance Measurement Mode (please refer to chapter ‘Setup Measurement
Mode’ for details),
when measurement starts, meter will automatically measure, calculate and
display measured result. In case measurement is in accordance with set balance condition, this
measurement finishes! At this time, you can press ‘STORE’(F1) to store current measurement
result. If USB is connected with computer, meter will print current measurement data.
Press ‘CANCEL’ key to end the measurement.
a) pH Electrode Calibration
Note:
1. Before each measurement, suggest re-calibrate the electrode. When calibration starts,
previous calibration data will be recovered.
2. Electrode should be re-calibrated after a period of using
3. Meter will remind automatically when the electrode calibration time interval is set to
suitable value.
pH measurement mode is shown as the chart.
On upper half of the screen is current measurement
data, meter displays current pH value (slope is set as
100.00%), potential value and temperature value.
At the bottom is current ‘Calibrated Result’.
AUTO:
Automatic distinguish function
MANUAL:
Manual distinguish function
GROUP:
Standard buffer solution group