MAINTENANCE AND TROUBLSHOOTING
62
Transmission Troubleshooting Chart
(continued)
LCD Message
Problem
Correction
CAN’T CONNECT TO REMOTE
LINK
A link to the access point was
established, but the link to the
destination failed.
Ensure the IP address is correct. If
problem persists, contact Mortara
Technical Service.
TIME SYNC FAULT
Possible incorrect version of
ELI Link
Install latest version.
UNABLE TO SAVE ORDER
Order storage failed.
Attempt to retransmit orders.
UNABLE TO SAVE WORK ITEM
DICOM order storage failed.
Directory full; mark records for deletion
or delete records.
INCORRECT RESPONSE
Connection established, then
failed.
Connection started but failed; attempt
to reconnect.
NO CUSTOM ID
Received orders failed.
Previous Custom ID not compatible
with current Custom ID, or no Custom
ID.
PAPER QUEUE FAULT
Unable to print.
Paper queue mark not
detected as expected.
Add paper; manually advance page
evenly past closure point of writer and
close writer cover and press STOP.
CONNECTION FAILED
Unable to transmit or receive
ECGs.
Check for correct baud rate, phone
number, and cable connections or site
number.
None
File not successfully
transmitted via LAN.
Check share permissions on host
device.
None
Unable to connect with LAN
with crossover cable.
Implement hub vs. crossover cable.
Disabled
Pressing SYNC key
Enable SYNC MODE and/or set SYNC
MEDIA in configuration